• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Disertación de Maestría
DOI
https://doi.org/10.11606/D.85.2016.tde-22022016-105213
Documento
Autor
Nombre completo
Camila Trindade de Oliveira
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2015
Director
Tribunal
Potiens, Maria da Penha Albuquerque (Presidente)
Nogueira, Maria do Socorro
Vivolo, Vitor
Título en portugués
Desenvolvimento de uma metodologia para calibração de câmaras de ionização de placas paralelas em feixes de raios X de energia baixa em termos de dose absorvida em água
Palabras clave en portugués
calibração
câmaras de ionização
dose absorvida
metrologia
raios X
Resumen en portugués
O objetivo deste trabalho foi estabelecer uma metodologia de calibração em termos de dose absorvida na água para câmaras de ionização de placas paralelas a serem utilizadas em feixes de radiação X de energia baixa (10 kV a 100 kV) para fins terapêuticos. Atualmente, no Brasil nenhum laboratório de calibração está oferecendo este tipo de serviço. Para o estabelecimento desta metodologia de calibração foram utilizadas duas câmaras de ionização PTW modelos 23344. Ambas as câmaras foram caracterizadas e testadas qualitativamente segundo recomendações de normas internacionais. Os testes de caracterização realizados foram: a medição da taxa de dose mínima efetiva, a curva de saturação, a eficiência da coleção de íons, a resposta quanto ao efeito da polaridade e a linearidade da câmara com a resposta. Os testes de controle de qualidade aplicados foram: estabilidade em curto e longo prazos, fuga de corrente sem irradiação, fuga de corrente após-irradiação e tempo de estabilização. Para a implantação da metodologia, foram estabelecidas as qualidades de radiação de T-10 a T-100, seguindo as recomendações do código de prática da AIEA. Desta forma, foi possível determinar a grandeza de referência, dose absorvida na água, utilizando-se as duas câmaras de ionizações e em todas as qualidades estabelecidas.
Título en inglés
Development of a methodology for calibration of parallel plate ionization chambers for X-ray beams of low energy in terms of absorbed dose to water
Palabras clave en inglés
absorbed dose
calibration
ionization chambers
metrology
X-ray
Resumen en inglés
The objective of this study was to establish a calibration methodology in terms of absorbed dose to water for parallel plate ionization chambers to be used in low energy X rays beams (10kV to 100kV) for therapy purposes. Currently, in Brazil no calibration laboratory is offering this type of service. In order to establish this calibration methodology was used two ionization chambers PTW models 23344. Both chambers were characterized and tested qualitatively according to recommendations of international standards. The characterization tests performed were: measuring the effective minimum dose rate, the saturation curve, the efficiency of ions collection, polarity effect and the linearity of the ionization chamber with the response. The quality control tests applied were: stability in the short and long term, leakage current without irradiation, leakage current after irradiation and settling time. For the implementation of the methodology, the radiation qualities the T- 10 to T-100 were established, following the recommendations of the IAEA code of practice. Thus, it was possible to determine the magnitude of reference absorbed dose in water, using the two ionization chambers in the all established qualities.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2016-03-04
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.