• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Master's Dissertation
DOI
https://doi.org/10.11606/D.85.2016.tde-22022016-105213
Document
Author
Full name
Camila Trindade de Oliveira
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2015
Supervisor
Committee
Potiens, Maria da Penha Albuquerque (President)
Nogueira, Maria do Socorro
Vivolo, Vitor
Title in Portuguese
Desenvolvimento de uma metodologia para calibração de câmaras de ionização de placas paralelas em feixes de raios X de energia baixa em termos de dose absorvida em água
Keywords in Portuguese
calibração
câmaras de ionização
dose absorvida
metrologia
raios X
Abstract in Portuguese
O objetivo deste trabalho foi estabelecer uma metodologia de calibração em termos de dose absorvida na água para câmaras de ionização de placas paralelas a serem utilizadas em feixes de radiação X de energia baixa (10 kV a 100 kV) para fins terapêuticos. Atualmente, no Brasil nenhum laboratório de calibração está oferecendo este tipo de serviço. Para o estabelecimento desta metodologia de calibração foram utilizadas duas câmaras de ionização PTW modelos 23344. Ambas as câmaras foram caracterizadas e testadas qualitativamente segundo recomendações de normas internacionais. Os testes de caracterização realizados foram: a medição da taxa de dose mínima efetiva, a curva de saturação, a eficiência da coleção de íons, a resposta quanto ao efeito da polaridade e a linearidade da câmara com a resposta. Os testes de controle de qualidade aplicados foram: estabilidade em curto e longo prazos, fuga de corrente sem irradiação, fuga de corrente após-irradiação e tempo de estabilização. Para a implantação da metodologia, foram estabelecidas as qualidades de radiação de T-10 a T-100, seguindo as recomendações do código de prática da AIEA. Desta forma, foi possível determinar a grandeza de referência, dose absorvida na água, utilizando-se as duas câmaras de ionizações e em todas as qualidades estabelecidas.
Title in English
Development of a methodology for calibration of parallel plate ionization chambers for X-ray beams of low energy in terms of absorbed dose to water
Keywords in English
absorbed dose
calibration
ionization chambers
metrology
X-ray
Abstract in English
The objective of this study was to establish a calibration methodology in terms of absorbed dose to water for parallel plate ionization chambers to be used in low energy X rays beams (10kV to 100kV) for therapy purposes. Currently, in Brazil no calibration laboratory is offering this type of service. In order to establish this calibration methodology was used two ionization chambers PTW models 23344. Both chambers were characterized and tested qualitatively according to recommendations of international standards. The characterization tests performed were: measuring the effective minimum dose rate, the saturation curve, the efficiency of ions collection, polarity effect and the linearity of the ionization chamber with the response. The quality control tests applied were: stability in the short and long term, leakage current without irradiation, leakage current after irradiation and settling time. For the implementation of the methodology, the radiation qualities the T- 10 to T-100 were established, following the recommendations of the IAEA code of practice. Thus, it was possible to determine the magnitude of reference absorbed dose in water, using the two ionization chambers in the all established qualities.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2016-03-04
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.