Disertación de Maestría
DOI
https://doi.org/10.11606/D.85.2019.tde-21112019-165533
Documento
Autor
Nombre completo
Daniel Litvac
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2019
Director
Tribunal
Caldas, Linda Viola Ehlin (Presidente)
Chubaci, Jose Fernando Diniz
Pinto, Teresa Cristina Nathan Outeiro
Título en portugués
Desenvolvimento de um kit postal dosimétrico para calibração de aplicadores clínicos de betaterapia no Brasil
Palabras clave en portugués
aplicadores clínicos
betaterapia
dosimetria
radiação beta
Resumen en portugués
O sistema de saúde no Brasil, como um todo, possui déficits em mão de obra e em infraestrutura. A betaterapia é uma das alternativas que o Sistema Único de Saúde e serviços particulares implantam para suprir a demanda. Aplicadores clínicos de 90Sr/90Y são utilizados para tratamentos dermatológicos e oftálmicos, substituindo o uso de aceleradores lineares, porque são de custo mais baixo, mais práticos e possuem 28,8 anos de meia-vida. A calibração e a recalibração periódica destes aplicadores, para verificação da taxa de dose absorvida, são essenciais para garantir a segurança nos tratamentos clínicos. Trabalhos de pesquisas mostram discrepâncias entre a calibração do fabricante e a realizada em laboratórios, podendo ter consequências danosas para os pacientes. Além disso, são poucos os laboratórios de calibração do tipo padrão primário para fontes de radiação beta no mundo. Na América Latina não existem laboratórios de calibração com este perfil. Foi desenvolvido neste trabalho um kit postal dosimétrico que permitiu a recalibração dos aplicadores clínicos utilizados em hospitais, de maneira prática, rápida e com instruções, por meio de seu envio aos proprietários, e depois retornando ao IPEN para posterior avaliação. Para isto, alguns materiais dosimétricos foram caracterizados em feixes padronizados de radiação beta, utilizando as técnicas de termoluminescência e luminescência opticamente estimulada para comparação de resultados da resposta quanto à reprodutibilidade, linearidade, curva dose-resposta e dependência angular, definindo as condições mais adequadas e o grupo de dosímetros para calibração dos aplicadores clínicos.
Título en inglés
Development of a dosimetric postal kit for calibration of clinical beta therapy applicators in Brazil
Palabras clave en inglés
beta radiation
beta therapy
clinical beta therapy applicators
dosimetry
Resumen en inglés
The Brazilian health system, as a whole, has a lack in manpower and infrastructure. In the radiotherapy area it is not different, the lack of equipment, as linear accelerators, instigates the search for alternatives that may replace this kind of instruments. Beta therapy is one of these alternatives that the Brazilian Unified Health System and private services implement to supply the demand. Clinical applicators with 90Sr/90Y sources are still used for dermatological and ophthalmic treatments, replacing the use of modern technologies in these procedures, because they are of lower cost, more practical, and they present 28.8 years of half-life. Calibration and periodic recalibration of these applicators, to verify the absorbed dose rates, are essential procedures for the quality assurance in clinical treatments. Research studies have shown discrepancies among the manufacturer calibrations and those performed in research laboratories, and this fact may lead to harmful consequences for patients. In addition, there are only a few primary standard calibration laboratories for beta radiation sources in the world. In Latin America there is no such calibration laboratory with this profile. A dosimetric postal kit was developed in this work which allowed the practical, quick and instructed recalibration of clinical applicators used in hospitals by sending them to the owners, and then returning to IPEN for further evaluation. Therefore, to meet the needs, dosimetric insulating materials were characterized in standard beta radiation beams, using a beta secondary standard system and using the technique of thermoluminescence and optically stimulated luminescence to compare respond results regarding reproducibility, linearity, dose-response curve and angular dependence, defining the most appropriate conditions and the dosimeter group for calibration of clinical applicators.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2019-12-03
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.