Thèse de Doctorat
DOI
https://doi.org/10.11606/T.85.2017.tde-20092017-155553
Document
Auteur
Nom complet
Felipe Belonsi de Cintra
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2016
Directeur
Jury
Caldas, Linda Viola Ehlin (Président)
Medeiros, Regina Bitelli
Moralles, Mauricio
Pinto, Teresa Cristina Nathan Outeiro
Vivolo, Vitor
Titre en portugais
Desenvolvimento de um calorímetro com núcleo de água e de uma unidade de tratamento de líquidos para dosimetria de radiação gama
Mots-clés en portugais
calorímetro de água
calorímetros
dosimetria de radiação gama
Resumé en portugais
Este trabalho consiste no desenvolvimento de um detector composto por um calorímetro com núcleo de água para dosimetria de feixes de 60Co e 137Cs com média intensidade. Além do calorímetro, foi também dimensionada a instrumentação, a metodologia de calibração e a medição de temperatura. O trabalho foi dividido em 4 etapas distintas: escolha do melhor projeto como base para desenvolvimento do calorímetro, dimensionamento de parâmetros e ajuste fino do projeto, construção e calibração. Durante o desenvolvimento deste trabalho foi construída uma Unidade de Tratamento de Água que permite a produção da água que o calorímetro emprega. Em seguida foi construído o calorímetro em sí com os conceitos das etapas anteriores. O equipamento foi testado tanto nas instalações da GMR quanto no CTR do IPEN, obtendo taxas de dose na água com fontes de 137Cs e 60Co com diversas atividades (3,3 TBq, 15,54 TBq e 7 TBq) e em algumas distâncias. Foram obtidas taxas de dose que variavam de 2 mGy/s até 15 mGy/s, dependendo do tipo de fonte e SDD escolhida. O projeto contou com simulações de códigos como MCNP5 e FLUENT 14 e foram essenciais na construção do equipamento. O objetivo deste trabalho foi aprimorar a infraestrutura do LCI, desenvolvendo um detector que deverá aumentar a gama dos serviços prestados pelo LCI com um tipo de medição dosimétrica ainda inexistente no país.
Titre en anglais
Development of a water calorimeter and a liquid treatment unit for gamma radiation dosimetry
Mots-clés en anglais
calorimeters
gamma dosimetry
water calorimeter
Resumé en anglais
This work aims to develop a radiation detector composed by a water calorimeter for beam dosimetry of 60Co and 137Cs gamma radiation with medium intensity. Besides the calorimeter itself, all the instrumentation, methodology, calibration and testing protocols were established. This work was divided into 4 parts: the choice of the best design; simulation and setting of the ideal operation parameters; adaptation of the project; construction and calibration. During the development of this work a water treatment unit was built which allows the production of distilled water for the calorimeter. Then the calorimeter itself was built with the concepts from the previous steps. The equipment has been tested both in the facilities of GMR as the CTR at IPEN, obtaining dose rates ( from 2 mGy/s up to 15 mGy/s) in water for 137Cs and 60Co sources in some activities (3,3 TBq, 15,54 TBq e 7 TBq) and distances. The MCNP5 and FLUENT 14 have been used and proved essential in the development of the prototype. The main goal was to improve the LCI infrastructure with a new radiation detector that will increase the calibration laboratory service range with a still absent kind of dosimetric measurement in Brazil.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2017-11-06
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.