• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Tesis Doctoral
DOI
https://doi.org/10.11606/T.85.2013.tde-19122013-161529
Documento
Autor
Nombre completo
Danilo Cardenuto Ferreira
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2013
Director
Tribunal
Bueno, Carmen Cecília (Presidente)
Barbosa, Renata de Faria
Caldas, Linda Viola Ehlin
Pascholati, Paulo Reginaldo
Potiens, Maria da Penha Albuquerque
Título en portugués
Desenvolvimento e calibração de um sistema dosimétrico de rotina em processamento por irradiação
Palabras clave en portugués
cobalto-60
dosimetria
fotodiodo
irradiador
Resumen en portugués
O desenvolvimento e calibração de um sistema dosimétrico de rotina baseado no fotodiodo comercial e de baixo custo (SFH 206) é apresentado neste trabalho. A sonda dosimétrica foi projetada para operar sem tensão de polarização no modo de corrente direta. As radiações foram realizadas no irradiador Panorâmico de Cobalto-60 no intervalo de taxas de dose de 8,1 Gy/h a 125 Gy/h. As fotocorrentes geradas no diodo, para cada taxa de dose, foram registradas utilizando um eletrômetro digital e foram armazenadas durante todo o tempo de exposição. A resposta em corrente do diodo foi medida em função do tempo de exposição em intervalos de 1 Gy até 200 Gy para doses acumuladas de até 15 kGy. Neste intervalo, a resposta em dose do diodo, em função da carga gerada pela dose, foi linear com coeficiente de correlação melhor que 0,998. Estes resultados foram comparados com os obtidos com filmes Gafchromic frequentemente utilizados em dosimetria de rotina. Para monitorar possíveis danos de radiação gama produzidos no diodo, as sensibilidades em corrente e em carga foram medidas em função da dose absorvida. Para doses até 15 kGy, não foi observado qualquer dano de radiação o que confirma a boa reprodutibilidade de resposta do diodo, melhor do que 3 %. Finalmente, devido aos pequenos erros experimentais ( 5% ) e boa resolução espacial do diodo, foi possível medir a dose de trânsito devida ao movimento da fonte radioativa de Cobalto-60 bem como a distribuição da taxa de dose no Irradiador Panorâmico.
Título en inglés
Development and calibration of a routine dosimetry system for radiation processing
Palabras clave en inglés
cobalt-60
dosimetry
irradiator
photodiode
Resumen en inglés
The development and calibration of a routine dosimetry system based on commercial, low cost photodiode (SFH 206) are presented in this work. The dosimeter probe was designed to operate unbiased in the direct current mode. The radiations were performed with Cobalt-60 Panoramic Irradiator facility in the dose-rate range of 8.1 Gy/h - 125 Gy/h. The photocurrents generated in the device, in each dose-rate, were registered with a digital electrometer and stored during the exposure time. The current response of the diode was measured as a function of the time in steps from 1 Gy up to 200 Gy with accumulated dose up to 15 kGy. In this range, the dose-response of the diode, given by the charge as function of dose, was linear with correlation coefficient better than 0.998. These results were compared with those obtained using Gafchromic film dosimetry often used in routine. To monitor possible gamma radiation effects produced on the diode, the current and charge sensitivities were measured as a function of the absorbed dose. For doses up to 15 kGy, it was not observed any radiation damage what confirms the reproducibility of the diode response better than 3 %. Finally, due to the small experimental errors ( 5% ) and good spatial resolution of the diode it was possible to measure the transit dose due to the movement of the Cobalt-60 radioactive source as well the dose-rate mapping in the Panoramic Irradiator.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2014-01-14
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.