Doctoral Thesis
DOI
https://doi.org/10.11606/T.85.2015.tde-19082015-093444
Document
Author
Full name
Neuza Taeko Okasaki Fukumori
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2015
Supervisor
Committee
Matsuda, Margareth Mie Nakamura (President)
Calvo, Wilson Aparecido Parejo
Pinto, Terezinha de Jesus Andreoli
Silva, Constancia Pagano Goncalves da
Suzuki, Katia Noriko
Title in Portuguese
Estudo da influência da atividade radioativa presente no gerador de 99Mo/99mTc na esterilidade do produto terminado
Keywords in Portuguese
efeito da radiação ionizante
esterilidade
esterilização
gerador de 99Mo/99mTc
microrganismos
Abstract in Portuguese
No processo de fabricação do gerador de 99Mo/99mTc é fundamental, não só evitar a introdução de contaminantes microbianos, como também avaliar o efeito que a atividade radioativa pode provocar em microrganismos. Os objetivos deste trabalho foram: estudar os processos de esterilização por radiação gama (60Co) dos acessórios e da coluna de alumina (Al2O3) que compõem o sistema do gerador de 99Mo/99mTc, e avaliar a eficácia da radiação gama na morte de microrganismos em geradores de 99Mo/99mTc na mínima (9,25 GBq) e na máxima (74 GBq) atividade radioativa. Acessórios do gerador foram irradiados com doses absorvidas de 15, 25 e 50 kGy e as colunas de alumina com 10, 15, 25 e 50 kGy. A alumina das colunas irradiadas foi avaliada por análise combinada de microscopia eletrônica de varredura (MEV) e análise elementar por espectroscopia de energia dispersiva (EDS), e difração de raios X. O Teflon® foi avaliado por termogravimetria (TGA) e calorimetria exploratória diferencial (DSC). Cálculos dosimétricos foram realizados para o alvo de maior dose na coluna do gerador de 9,25 GBq e 74 GBq. Suspensões quantificadas de B. subtilis e esporos de B. pumilus foram inoculadas em colunas irradiadas a 25 kGy e observou-se que a coluna de alumina representou uma barreira física para a contaminação do eluato, pois a recuperação microbiana foi reduzida. Verificou-se que o B. subtilis na forma vegetativa mostrou menor afinidade pela alumina em relação aos esporos de B. pumilus. A recuperação microbiana foi menor no gerador com atividade radioativa em relação ao gerador sem radioatividade. Foi desenvolvido um indicador biológico de esporos de B. pumilus que consistiu em uma tira de alumínio com sílica gel colocado externamente à coluna e as quantidades recuperadas tanto no gerador de 9,25 GBq quanto no de 74 GBq mostraram redução na viabilidade. O uso de filtros Sep-Pak® e 0,22 μm no sistema do gerador mostrou eficiência na retenção de 4,9 x 106 UFC de B. subtilis, garantindo a esterilidade do produto terminado.
Title in English
Study of the influence of 99Mo/99mTc generator radioactivity on the sterility of the finished product
Keywords in English
99Mo/99mTc generator
ionizing radiation effect
microorganisms
sterility
sterilization
Abstract in English
In the manufacturing process of the 99Mo/99mTc generator it is essential not only to prevent the introduction of microbial contaminants, but also evaluate the effect of the radioactivity in microorganisms. The objectives of this work were to study the sterilization processes by gamma rays (60Co) of accessories and alumina column (Al2O3) that take part in the generator system and assess the effectiveness of gamma radiation in the death of microorganisms in 99Mo/99mTc generators of minimum (9.25 GBq) and maximum (74 GBq) radioactivities. Accessories were irradiated with 15, 25 and 50 kGy absorbed doses and alumina columns with 10, 15, 25 and 50 kGy. Alumina from irradiated columns were examined by combined analysis of scanning electron microscopy (SEM) and elemental analysis by energy dispersive spectroscopy (EDS), and X-ray diffraction. Teflon® was evaluated by thermogravimetric analysis (TGA) and differential scanning calorimetry (DSC). Dosimetric calculations were made for the highest dose target on 9.25 GBq and 74 GBq generator columns. Quantified suspensions of B. subtilis and B. pumilus spores were inoculated into columns irradiated at 25 kGy and it was observed that the alumina column represented a physical barrier to contamination of the eluate, as microbial recovery was reduced. It was found that B. subtilis in vegetative form showed lower affinity for alumina when compared to B. pumilus spores. Microbial recovery was lower in the generator with radioactivity than with the generator without radioactivity. A biological indicator of B. pumilus spores was developed which consisted of an aluminum strip with silica gel placed externally to the column and the microbial recovery in both 9.25 and 74 GBq generators showed reduction in viability. The use of Sep-Pak® and 0.22 μm filters in the generator system has shown efficiency in the retention of 4.9 x 106 CFU of B. subtilis, ensuring the sterility of the finished product.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2015-08-19
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.