Disertación de Maestría
DOI
https://doi.org/10.11606/D.85.2009.tde-16112009-153552
Documento
Autor
Nombre completo
Patrícia de Lara Antonio
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2009
Director
Tribunal
Caldas, Linda Viola Ehlin (Presidente)
Nersissian, Denise Yanikian
Vivolo, Vitor
Título en portugués
Estudo comparativo entre métodos de calibração de aplicadores clínicos de radiação beta
Palabras clave en portugués
Aplicadores clínicos
Calibração
Radiação Beta.
Resumen en portugués
Os aplicadores clínicos de 90Sr+90Y são instrumentos utilizados em procedimentos de braquiterapia e devem ser calibrados periodicamente, de acordo com normas e recomendações internacionais. Neste trabalho, foram estudados quatro métodos de calibração de aplicadores dermatológicos e oftálmicos, comparando-se os resultados com os fornecidos pelos certificados de calibração dos fabricantes dos aplicadores. Os métodos envolveram o aplicador clínico padrão do Laboratório de Calibração de Instrumentos (LCI), calibrado pelo laboratório padrão primário americano do National Institute of Standards and Technology, como referência; um aplicador da Amersham, também pertencente ao LCI, como referência; uma mini-câmara de extrapolação desenvolvida no IPEN como padrão absoluto; e dosimetria termoluminescente. A mini-câmara de extrapolação e uma câmara de extrapolação comercial PTW foram estudadas com relação ao seu desempenho por meio de testes de controle de qualidade, como corrente de fuga, repetitividade e reprodutibilidade. A distribuição de dose em profundidade na água, estudo de grande importância na dosimetria de aplicadores clínicos, foi determinada utilizando-se a mini-câmara de extrapolação e dosímetros termoluminescentes. Os resultados obtidos foram considerados satisfatórios para os dois casos, quando comparados com os dados fornecidos pela norma IAEA (2002). Além disso, foi desenvolvido um sistema postal dosimétrico para a calibração de aplicadores clínicos por meio da técnica da termoluminescência, para ser enviado para clínicas e hospitais, sem a necessidade do transporte das fontes ao LCI do IPEN para calibração.
Título en inglés
COMPARATIVE STUDY AMONG CALIBRATION METHODS OF CLINICAL APPLICATORS OF BETA RADIATION
Palabras clave en inglés
Aplicadores clínicos
Calibração
Radiação beta
Resumen en inglés
90Sr+90Y clinical applicators are instruments used in brachytherapy procedures and they have to be periodically calibrated, according to international standards and recommendations. In this work, four calibration methods of dermatological and ophthalmic applicators were studied, comparing the results with those given by the calibration certificates of the manufacturers. The methods included the use of the standard applicator of the Calibration Laboratory (LCI), calibrated by the National Institute of Standards and Technology; an Amersham applicator (LCI) as reference; a mini-extrapolation chamber developed at LCI as an absolute standard; and thermoluminescent dosimetry. The mini-extrapolation chamber and a PTW commercial extrapolation chamber were studied in relation to their performance through quality control tests of their response, as leakage current, repetibility and reproducibility. The distribution of the depth dose in water, that presents high importance in dosimetry of clinical applicators, was determined using the miniextrapolation chamber and the thermoluminescent dosimeters. The results obtained were considered satisfactory for the both cases, and comparable to the data of the IAEA (2002) standard. Furthermore, a dosimetry postal kit was developed for the calibration of clinical applicators using the thermoluminescent technique, to be sent to clinics and hospitals, without the need of the transport of the sources to IPEN for calibration.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2009-11-27
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.