Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.85.2007.tde-16052012-110041
Document
Auteur
Nom complet
Erick Yukio Mizuno
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2007
Directeur
Jury
Rodrigues, Letícia Lucente Campos (Président)
Ludwig, Zélia Maria da Costa
Rodrigues Junior, Orlando
Titre en portugais
Desenvolvimento e caracterização de um gel de alanina para aplicação na medida da distribuição da dose de radiação usando a técnica de espectrofotometria
Mots-clés en portugais
alanina
dosímetro gel
espectrofotometria
Resumé en portugais
Este trabalho tem por objetivo a composição e caracterização de um dosímetro à base de DL-Alanina e Ferro-ll em forma de gel para aplicações em dosimetría para campos de doses da radiação gama do Co-60 experimentadas em procedimentos de radioterapia. O aminoácido DL-Alanina é dissolvido em solução ácida contendo o Fe-ll e adicionado o gel. Procedeu-se a uma série de ensaios variando reagente a reagente e realizando procedimentos para avaliar o comportamento em função do tempo do gel dosimétrico obtido. Os espectros de absorção antes e após a irradiação apresentam máximos em 457 nm e 588 nm, respectivamente. Esses dois picos correspondem aos comprimentos de onda de absorção das espécies Fe-ll e Fe-lll, respectivamente; estando ambas as espécies em equilíbrio. Com a irradiação, ocorre a formação de radicais que oxidam o Fe-ll em Fe-lll, alterando com isso o balanço entre as duas espécies - daí o aumento da concentração de Fe-lll - fazendo com que mude a tonalidade do gel, sendo, portanto, possível de ser determinada a dose absorvida. Nas análises dos espectros e respectivas curvas dose-resposta observou-se que o gel dosimétrico apresenta linearidade no intervalo de doses entre 0,5 e 40 Gy. A partir dos resultados obtidos de linearidade, estabilidade e menor dose detectável - 0,3 Gy - conclui-se que o presente Gel Dosimétrico à base de DL-Alanina e Fe-ll apresenta excelente potencial para a finalidade que é proposto, qual seja a determinação do campo de doses em radioterapia e futura aplicação como padrão na determinação desses campos, em 3D, utilizando-se a técnica de Imagem por Ressonância Magnética - IRM.
Titre en anglais
Developement and characterization of an DL-Alanine gel to be applied in the measurement of the dose distribution with the spectrophotometry technique
Mots-clés en anglais
alamines
cobalt 60
dosimeters
gamma radiation
gels
magnetic resonance
radiation dose distributions
radiotherapy
spectrothotometry
Resumé en anglais
The aim of this work is to develop a DL-Alanine, Fe-ll based gel dosimeter to be applied in the dosimetry of the Co-60 gamma-radiation fields in the dose range of radiation therapy procedures. The aminoacid DL-Alanine is dissolved in an acid solution containing the Fe-ll and added to the gel. A series of essays was performed with different chemicals and different procedures were made to evaluate the behavior of the obtained dosimetric gel as function of the time. The absorption spectra, before and after the irradiation, present maximum in 457 nm and 588 nm, respectively. These two peaks correspond to absorption wavelengths of the chemical species Fe-ll and Fe-lll, respectively, both species being in equilibrium. With the irradiation, the formation of the radicals that oxide the Fe-ll into Fe-lll occurs, altering the chemical balance of both species - hence the increase of the Fe-lll - and causing the gel to change its tonality, in such a way that it is possible to determine the absorved dose. In the analysis of the spectra and respective dose-response curves it was observed that the dosimetric gel shows linearity in dose range of 0.5 to 40 Gy. From the obtained linearity results, stability and lower detectable dose - 0,3 Gy - it is possible to conclude that the present DL-Alanine, Fe-ll based dosimetric gel presents an excellent potential to the application to which it is proposed, namely, the determination of the dose fields in radiation therapy and its future application as a standard in the determination of these fields, in 3D, using the Magnetic Resonance Imaging.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2012-05-30
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.