Tese de Doutorado
DOI
https://doi.org/10.11606/T.85.2007.tde-16052012-104232
Documento
Autor
Nome completo
Pedro Pereira Rodrigues
E-mail
Unidade da USP
Área do Conhecimento
Data de Defesa
Imprenta
São Paulo, 2007
Orientador
Banca examinadora
Rodrigues, Letícia Lucente Campos (Presidente)
Caldas, Linda Viola Ehlin
Costa, Paulo Roberto
Ghilardi Netto, Thomaz
Hamada, Margarida Mizue
Título em português
Estudos microdosimétricos usando um sistema de irradiação de nêutrons rápidos filtrados de reator de pesquisa para aplicação de radiobiologia
Palavras-chave em português
dosimetria de nêutrons
microdosimetria
Resumo em português
Neste trabalho, medidas microdosimétricas foram realizadas usando um contador proporcional equivalente à tecido - TEPC. com uma cavidade esférica de diâmetro de 1.27 cm. O TEPC foi preenchido com gás propane puro, C3HS, à pressão de 5,6 kPa (42 Torr), que é equivalente a 1,3 μm em diâmetro de unidade de densidade do tecido. O instrumento de medida microdosimétrica foi irradiado com radiação de nêutrons rápidos do reator de pequisa do Nuclear Science Center da Texas A&M University, em College Station,-Texas. Os feixes de nêutrons rápidos foram emitidos com três diferentes valores de potência. 0,5, 1,0, e 2,0 kVV, durante 1 hora para alto ganho e o mesmo tempo para baixo ganho, totalizando 2 horas para cada potência, com 0,0083 Gy/min de taxa de dose. O neutron foi filtrado usando o sistema de irradiação de néutrons rápidos fortemente nitrados (FNIS). do Nuclear Science Center, para obter uma redução da contaminação da radiação de neutron por radiação gama e assim obter espectros microdosimetricos de neutrons como, distribuição de freqüência de energia lineal e distribuição de dose de energia lineal, com boa precisão, e outras grandezas como, freqüência média de energia lineal, dose média de energia lineal, dose absorvida, dose equivalente e fator de qualidade médio de neutron rápido. Os resultados obtidos foram satisfatórios, com os espectros microdosimetricos de nêutrons mostrando uma contaminação de radiação gama abaixo de 5 %, especialmente para distribuição de dose de energia lineal. Os resultados obtidos neste trabalho foram comparados com outros da literatura científica, que usaram outros procedimentos para a redução da contaminação do neutron por radiação gama. estando em concordância com eles.
Título em inglês
Microdosimetric studies using a filtered fast neutron irradiation system of research reactor to application in radiation biology
Palavras-chave em inglês
fast neutrons
filters
microdosimetry
radiations
radiobiology
research reactors
wall-less counters
Resumo em inglês
In this work, microdosimetric measurements were performed using a Wall-less Tissue Equivalent Proportional Counter - TEPC was filled with spherical cavity with an inner diameter of 1.27 cm. The TEPC was tilled with pure propane gas, C3H8 at 5.6 kPa (42 Torr) pressure, which is equi\alent to 1.3μm in diameter of unit density tissue. The microdosimetric measurement device was irradiated with fast neutron radiation from Texas A&M University Nuclear Science Center research reactor, in College Station, Texas. The fast neutron beams were emitted with three different power values, 0.5, 1.0 and 2.0 kW. during Ih for both high gain and low gain, totalizing two hours for each power with 0.0083 Gy/min of dose rate. The neutron was filtered using the heavily filtered fast neutron irradiation s\stem (FNIS). from Nuclear Science Center, to obtain a decrease of neutron radiation contamination by gamma ray and so, to gain the neutron microdosimetric spectra as. frequency distribution of lineal energy, dose distribution of lineal energy with good precision, and another quantities as frequency-mean of lineal energy, dose- mean of lineal energy, absorbed dose, equivalent dose and average quality factor of fast neutron. The obtained results were satisfactory, with the neutron microdosimetric spectra showing a gamma ray contamination under 5 %, especially to dose distribution of lineal energy. The results obtained in this work were in agreement when compared with another results from scientific literature, which used another procedure to reduce the neutron contamination by gamma ray.
 
AVISO - A consulta a este documento fica condicionada na aceitação das seguintes condições de uso:
Este trabalho é somente para uso privado de atividades de pesquisa e ensino. Não é autorizada sua reprodução para quaisquer fins lucrativos. Esta reserva de direitos abrange a todos os dados do documento bem como seu conteúdo. Na utilização ou citação de partes do documento é obrigatório mencionar nome da pessoa autora do trabalho.
Data de Publicação
2012-05-30
 
AVISO: Saiba o que são os trabalhos decorrentes clicando aqui.
Todos os direitos da tese/dissertação são de seus autores
CeTI-SC/STI
Biblioteca Digital de Teses e Dissertações da USP. Copyright © 2001-2024. Todos os direitos reservados.