Disertación de Maestría
DOI
https://doi.org/10.11606/D.85.2006.tde-16052012-103302
Documento
Autor
Nombre completo
Fernanda Cristina Salvador Soares
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2006
Director
Tribunal
Caldas, Linda Viola Ehlin (Presidente)
Nersissian, Denise Yanikian
Potiens, Maria da Penha Albuquerque
Título en portugués
Avaliação de dispositivos de proteção individual utilizados em radiologia diagnóstica
Palabras clave en portugués
avental plumbífero
blindagem
radioproteção
raios X
Resumen en portugués
Neste trabalho foram avaliados aventais de proteção e blindagens de ovário de cinco fabricantes segundo as normas NBR / IEC 61331-1 e NBR / IEC 61331-3 de 2004. Para a determinação da espessura equivalente de atenuação foram aplicadas três metodologias diferentes (interpolação linear, modelo de Archer e somatório de exponenciais), uma vez que a norma não indica como este valor deve ser obtido. Além disso, foram estimadas as incertezas associadas a cada método e a influência do número de medições na incerteza combinada. As características de confecção avaliadas foram: documento acompanhante, projeto, materiais, dimensões e marcação da etiqueta. Para esta avaliação foi elaborado um formulário que contemplava todas as exigências da NBR / IEC 61331-3. Os resultados mostraram uma grande diferença entre os valores de espessura equivalente de atenuação declarados e medidos. A comparação dos resultados das três metodologias aplicadas mostrou que há pouca variação entre os valores obtidos e entre as incertezas associadas nas diferentes metodologias. Foi possível observar também que o número de medições não contribui significativamente para o aumento da incerteza nas três metodologias. A metodologia que melhor se adequou à rotina do laboratório foi a metodologia da interpolação linear com cinco medições da taxa de kerma no ar. A discrepância entre os resultados apresentados neste trabalho e as exigências das normas estudadas mostra a necessidade de tornar compulsória a certificação dos dispositivos de proteção, contribuindo assim para o aumento da proteção radiológica do usuário.
Título en inglés
Evaluation of personal protective devices used in diagnostic radiology
Palabras clave en inglés
attenuation
data covariances
experimental data
ionization chambers
kerma
protective clothing
radiation protection
radiation quality
safety standards
thickness
X-ray equipment
Resumen en inglés
In this work protective devices of five manufacturers were evaluated according to the NBR/IEC 61331-1 and NBR/IEC 61331-3 standards. Three different methodologies (linear interpolation, Archer model and sum of exponentials) were applied for the determination of the attenuation equivalent, since the standard does not indicate how it must be determined. Moreover, the uncertainties associated to each method, and the influence of the number of measurements in the combined uncertainty were estimated. The evaluated manufacture characteristics were: accompanying document, design, materials, dimensions and label marking. For this evaluation a check list about the requirements of the NBR/IEC 61331-3 standard was elaborated. The results showed a great difference between nominal and measured attenuation equivalent values. The comparison of the results using the three methodologies showed small variations among the obtained values and among the associated uncertainties in the different methodologies. It was possible to observe that the number of measurements does not contribute significantly for the increase of the uncertainty in all three methodologies. The best methodology for the laboratory routine is the linear interpolation methodology, with five measurements for each air kerma rate value. The discrepancy between the results obtained in this work and the requirements of the applied standards show the need to adopt a compulsory certification process for protective devices, thus contributing for the increase of the radiation protection of the users.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2012-05-30
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.