Master's Dissertation
DOI
https://doi.org/10.11606/D.85.2014.tde-14112014-142426
Document
Author
Full name
Paulo Marcio de Oliveira
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2014
Supervisor
Committee
Rodrigues, Letícia Lucente Campos (President)
Rocha, Felícia Del Gallo
Vivolo, Vitor
Title in Portuguese
Avaliação de dose nos sistemas de irradiação de bolsas de sangue para esterilização
Keywords in Portuguese
dosimetria química
dosímetro Fricke Gel
irradiação de bolsas de sangue
irradiação de hemocomponentes do sangue
Abstract in Portuguese
O procedimento de esterilização do sangue é adotado para que sejam mínimas as condições necessárias de manifestação da doença do enxertoversus- hospedeiro associada à transfusão de sangue (DECHT), síndrome, geralmente fatal. Neste trabalho foi utilizada a solução Fricke gel desenvolvida no IPEN para a obtenção de um objeto simulador de bolsa de sangue empregada no processo de irradiação para esterilização de sangue em um irradiador Gammacell Elan. O objetivo desse trabalho é determinar a homogeneidade da dose aplicada no processo de esterilização de bolsas de sangue. Foram utilizadas as técnicas de Espectrofotometria de Absorção Óptica (EAO), Densitometria Óptica (DO), Termoluminescência (TL) e Ressonância Magnética por Imagem (RMI) e os softwares Numaris® versão 4 (Syngo-MR-A30) e MatLab® (versão 9.0) para a aquisição e reconstrução das imagens. O estudo foi realizado posicionando no canister do irradiador inicialmente uma bolsa e, posteriormente, duas bolsas e três bolsas e aplicando uma dose nominal de 25 Gy. O fabricante do irradiador Gammacell Elan recomenda que seja irradiado em cada procedimento o valor máximo de três bolsas de sangue. Os resultados obtidos indicam que o irradiador Gammacell Elan não demostrou homogeneidade na distribuição da dose. Irradiando uma bolsa de sangue foi obtida uma diferença de dose de 3,65 %, enquanto para duas bolsas de sangue essa diferença caiu para 2,65 % e quando posicionadas três bolsas de sangue a diferença de dose foi de 1,88 %. O tempo de abertura de fechamento do equipamento de irradiação é responsável pela diferença de dose 3,17 % acima da dose prevista à realização das irradiações de bolsas de sangue. Os resultados obtidos usando o filme radiocrômico Grafchromic MD-55 de dupla face aplicada à avaliação de dose no sistema de rotação de irradiação não apresentou resultados reproduzíveis.
Title in English
Evaluation of radiation dose in systems of blood bag for sterilization
Keywords in English
chemical dosimetry
Fricke Gel dosimeter
irradiation of blood bags
irradiation of blood components from the blood
Abstract in English
The blood sterilization procedure is adopted so that the necessary conditions of manifestation of graft-versus-host syndrome, disease associated with blood transfusion (AT-GVHD), usually a fatal disease, are minimal. In this work the Fricke gel solution developed in IPEN was used for obtaining a phantom of blood bag used in the irradiation process for sterilization of blood in a Gammacell irradiator Elan. The objective of this study was to evaluate the homogeneity of the applied radiation sterilization dose in the blood bags. Optical Absorption Spectroscopy (OAS), Optical Densitometry (OD), Thermoluminescence (TL) and Magnetic Resonance Imaging (MRI) technics and Numaris® software version 4 (Syngo-MR-A30) and MatLab® (version 9.0) softwares were used to images acquisition and reconstruction. The study was conducted positioning one bag in the canister irradiator initially and subsequently two and three bags together and applying a nominal dose of 25 Gy. The Elan Gammacell irradiator manufacturer recommends that the maximum values of three blood bags are irradiated in each procedure. The results indicate that the Gammacell irradiator Elan did not presents homogeneous dose distribution. Radiating a blood bag a dose difference of 3.65 % was obtained, while for two bags of blood that difference dropped to 2.65% and when positioned three blood bags the difference in dose was 1.88 %. The opening and closing times of the irradiation source is responsible for an increasing in the planned irradiation dose of 3.17%. The results obtained using radiochromic double-sided Grafchromic MD-55 film applied to dose evaluation in the rotation irradiation system did not present reproducible results.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2014-12-22
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.