Título en portugués
Desenvolvimento de uma metodologia de calibração e testes de medidores de produto Kerma-Área
Palabras clave en portugués
calibração
patient dose calibrator
produto kerma area
Resumen en portugués
A grandeza produto kerma-área (PKA) é importante para estabelecer níveis de referência em exames de radiologia diagnóstica. Essa grandeza pode ser obtida por meio de medidores do PKA. O uso desses medidores é fundamental para avaliar a dose de radiação em procedimentos radiológicos, além de ser um bom indicador para que os limites de dose na pele do paciente não sejam excedidos. Algumas vezes, esses medidores vêm acoplados a equipamentos de radiação X, o que dificulta sua calibração. Neste trabalho foi desenvolvida uma metodologia de calibração de medidores do PKA. O instrumento utilizado para este fim foi o Patient Dose Calibrator (PDC). Ele foi desenvolvido para ser utilizado como referência na verificação da calibração de medidores do PKA e kerma no ar usados na dosimetria de pacientes e para verificação da consistência e do comportamento de sistemas de controle de exposição automáticos. Por se tratar de um equipamento novo, que, no Brasil, ainda não é utilizado como equipamento de referência para calibração, foi realizado, também o controle de qualidade deste equipamento, com testes de caracterização, a calibração e a avaliação da dependência energética. Após os testes, ficou provado que o PDC pode ser utilizado como instrumento de referência para a calibração a ser realizada in situ, de forma que as características de cada equipamento de radiação X onde os medidores do PKA são utilizados sejam consideradas. A calibração foi, então, realizada com medidores do PKA portáteis e em um equipamento de radiologia intervencionista que possui um medidor do PKA acoplado. Os resultados foram bons e ficou provada a necessidade de calibração desses medidores e a importância da calibração in situ com um medidor de referência.
Título en inglés
Development of a calibration methodology and tests of kerma area product meters
Palabras clave en inglés
calibration
kerma area product
patient dose calibrator
Resumen en inglés
The quantity kerma area product (PKA) is important to establish reference levels in diagnostic radiology exams. This quantity can be obtained using a PKA meter. The use of such meters is essential to evaluate the radiation dose in radiological procedures and is a good indicator to make sure that the dose limit to the patient's skin doesnt exceed. Sometimes, these meters come fixed to X radiation equipment, which makes its calibration difficult. In this work, it was developed a methodology for calibration of PKA meters. The instrument used for this purpose was the Patient Dose Calibrator (PDC). It was developed to be used as a reference to check the calibration of PKA and air kerma meters that are used for dosimetry in patients and to verify the consistency and behavior of systems of automatic exposure control. Because it is a new equipment, which, in Brazil, is not yet used as reference equipment for calibration, it was also performed the quality control of this equipment with characterization tests, the calibration and an evaluation of the energy dependence. After the tests, it was proved that the PDC can be used as a reference instrument and that the calibration must be performed in situ, so that the characteristics of each X-ray equipment, where the PKA meters are used, are considered. The calibration was then performed with portable PKA meters and in an interventional radiology equipment that has a PKA meter fixed. The results were good and it was proved the need for calibration of these meters and the importance of in situ calibration with a reference meter.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.