Thèse de Doctorat
DOI
https://doi.org/10.11606/T.85.2018.tde-13032018-101043
Document
Auteur
Nom complet
Iremar Alves da Silva Junior
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2017
Directeur
Jury
Potiens, Maria da Penha Albuquerque (Président)
Bottaro, Márcio
Caldas, Linda Viola Ehlin
Silva, Teógenes Augusto da
Vivolo, Vitor
Titre en portugais
Desenvolvimento de uma metodologia de calibração de monitores de contaminação de superfície considerando o mapeamento da uniformidade das fontes extensas de referência
Mots-clés en portugais
calibração
fator de correção
fonte extensa de referência
MCNP
monitor de contaminação de superfície
uniformidade
Resumé en portugais
Uma das exigências na calibração de monitores de radiação de contaminação de superfície consiste no uso de fontes extensas de referência com um valor de uniformidade superior a 90%. Entretanto, foi verificado em alguns laboratórios de calibração que suas fontes não atendiam essa exigência, o que obrigaria esses laboratórios a não utilizarem tais fontes em suas calibrações. Foi neste contexto que foi desenvolvido um estudo para propor o uso dessas fontes, desde que sejam utilizados fatores de correção, que são propostos neste trabalho, que foram avaliados e desenvolvidos usando o código de transporte MCNP como ferramenta de análise e avaliação, a partir dos dados de mapeamento da uniformidade. Também foi desenvolvido um software para calcular esses fatores de correção, utilizando o mapeamento da uniformidade das fontes como dados de entrada e um sistema automatizado para calibração de monitores de radiação de contaminação de superfície. O grande ganho com esse trabalho foi poder utilizar fontes extensas de referência, mesmo quando essas estão em desacordo com as exigências de uniformidade da norma ABNT ISO 8769:2016.
Titre en anglais
Development of a calibration methodology of surface contamination monitors considering the uniformity mapping of the large area reference sources
Mots-clés en anglais
calibration
correction factor
large-area reference source
MCNP
surface contamination monitor
uniformity
Resumé en anglais
One of the requirements in the calibration of surface contamination radiation monitors is the use of wide area reference sources with a uniformity value greater than 90%. However, it has been found, in some calibration laboratories, that their sources were not meeting this requirement, which would not allow these laboratories to use such sources in their calibration procedures. It was in this context that a study was developed to propose the use of these sources, provided that correction factors are used, which are proposed in this work, which were evaluated and developed using the MCNP radiation transport code as an evaluation tool, from the uniformity mapping data. A software was also developed to calculate these correction factors using the source uniformity mapping as an input data along with an automated system for calibration of surface contamination radiation monitors. The great gain in this work was to be able to use wide area reference sources, even when they are in disagreement with the uniformity requirements of ISO 8769:2016.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2018-04-11
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.