Disertación de Maestría
DOI
https://doi.org/10.11606/D.85.2008.tde-10072009-095813
Documento
Autor
Nombre completo
Carlos Augusto Pires
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2008
Director
Tribunal
Koskinas, Marina Fallone (Presidente)
Camargo, Sonia Pompeu de
Moreira, Denise Simões
Título en portugués
Padronização dos Radionuclídeos 201Tl e 55Fe em sistema de coincidências 4(PC)-NaI(Tl)
Palabras clave en portugués
ferro-55
padronização
sistema de concidências
Tálio-201
Resumen en portugués
No presente trabalho foi desenvolvido o método de padronização de radionuclídeos produzidos em cíclotrons utilizando o sistema de coincidências 4(PC)-NaI(Tl) composto de um contador proporcional operando a 0,1 MPa. Os radionuclídeos selecionados, foram o 201Tl, de uso em medicina nuclear e o 55Fe padrão primário muito utilizado na calibração de espectrômetros de raios-x. O 201Tl desintegra pelo processo de captura eletrônica, seguido pela emissão de radiação gama. A taxa de desintegração foi determinada pela técnica de extrapolação linear da eficiência empregando dois métodos: discriminação eletrônica e absorvedores externos. As fonte radioativas foram preparadas em substratos de Collodion com espessura de 20 g cm-2. O sistema eletrônico utilizado é o convencional para este tipo de medida. Os eventos observados foram registrados empregando o método TAC. O 55Fe desintegra pelo processo de captura eletrônica decaindo diretamente para o estado fundamental do 54Mn, emitindo raios-x com aproximadamente 6 keV. A taxa de desintegração foi determinada pelo método do traçador. Esta técnica foi aplicada utilizando dois radionuclídeos traçadores, que decaem pelo processo de captura eletrônica seguido da emissão de fótons, são eles 51Cr e 54Mn. As medidas foram feitas com 1 e 2 absorvedores de Al com espessura de 150 g cm-2. A atividade foi obtida extrapolando para zero espessura de Al. As incertezas envolvidas e suas correlações foram analisadas seguindo a metodologia de matriz de covariâncias.
Título en inglés
STANDARDIZATION OF 201Tl AND 55Fe RADIONUCLIDES IN A 4(PC)-NaI(Tl) COINCIDENCE SYSTEM
Palabras clave en inglés
ferro-55
padronização
sistema de coincidências
Tálio-201
Resumen en inglés
In the present work the procedure for the standardization of radionuclides using the 4(PC)-NaI(Tl) coincidence system was developed. The radionuclides selected were 201Tl, used in nuclear medicine, and 55Fe primary standard source, used for x-ray spectrometers calibration. The 4(PC)-NaI(Tl) is composed of a 4 proportional counter operated at 0.1MPa coupled to two NaI(Tl) crystals. The 201Tl decays by electron capture process followed by a prompt gamma-ray. The disintegration rate was determined by extrapolation technique using two methods: electronic discrimination and external absorbers. The radioactive sources were prepared in a 20 g cm-2 thick Collodion film. The conventional electronic system was used. The observed events were registered by the TAC method. The 55Fe decays by electron capture process to the ground state of 55Mn, emitting x-rays with around 6 keV. The standardization was obtained by the tracing method. This technique was applied using two radionuclides, which decay by electron capture process followed by a prompt gamma-ray, namely 51Cr and 54Mn, as tracers. Measurements with 1 and 2 aluminum foils, each 150 g cm-2 thick were carried out. The activity was obtained by extrapolation for zero thickness Al foil. The uncertainties were treated by means of matrix covariance methodology and takes into account all correlations involved.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2009-07-16
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.