Disertación de Maestría
DOI
https://doi.org/10.11606/D.85.2017.tde-09122016-091446
Documento
Autor
Nombre completo
Maria Kuznetsova
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2016
Director
Tribunal
Koskinas, Marina Fallone (Presidente)
Maidana, Nora Lia
Potiens Junior, Ademar José
Título en portugués
Medida absoluta da atividade de 14C pelos métodos CIEMAT/NIST, TDCR e em sistema de coincidência 4πβ-γ
Palabras clave en portugués
cintilação líquida
emissores beta puros
medicina nuclear
Resumen en portugués
Neste trabalho foi padronizada uma solução de 14C emissor beta puro com energia máxima de 156 keV, por meio de três diferentes métodos: CIEMAT/NIST e TDCR (triple-to-double coincidence ratio) em sistemas de cintilação líquida e pelo método do traçador, em sistema de coincidências 4πβ-γ. O sistema de cintilação líquida TRICARB, equipado com dois tubos fotomultiplicadores, foi usado para aplicação do método CIEMAT/NIST, usando um padrão de 3H emissor beta puro com energia máxima de 18,7 keV como traçador de eficiência. O sistema de cintilação líquida HIDEX 300SL, equipado com três tubos fotomultiplicadores, foi utilizado para as medidas pelo método TDCR. As amostras de 14C e 3H, medidas nos sistemas de cintilação foram preparadas usando-se três coquetéis cintiladores comerciais Ultima Gold, Optiphase Hisafe3 e InstaGel-Plus a fim de comparar seu desempenho nas medidas.Todas as amostras foram preparadas com 15 mL de coquetel cintilador, em frascos de vidro com baixa concentração de potássio. Alíquotas conhecidas de solução radioativa foram depositadas nos coquetéis cintiladores. Para a variação do parâmetro indicador de quenching, foram utilizados: um carregador de nitro metano e 1 mL de água destilada. Para a padronização pelo método do traçador no sistema de coincidências 4πβ-γ, foi utilizado 60Co como emissor beta gama. As medidas foram feitas no sistema de coincidências por software SCS, usando discriminação eletrônica para alterar a eficiência beta. O comportamento da curva de extrapolação foi predito por meio do código Esquema, que utiliza a técnica de Monte Carlo. Os resultados da atividade da solução de 14C obtida pelos três métodos utilizados mostraram uma boa concordância dentro da incerteza experimental.
Título en inglés
Primary standardization of C-14 by means of CIEMAT/NIST, TDCR and 4πβ-γ methods
Palabras clave en inglés
beta pure emitters
liquid scintilation
nuclear medicine
Resumen en inglés
In this work, the primary standardization of 14C solution, which emits beta particles of maximum energy 156 keV, was made by means of three different methods: CIEMAT/NIST and TDCR (triple-to-double coincidence ratio) methods in liquid scintillation systems and the tracing method, in the 4πβ-γ coincidence system. TRICARB LSC (Liquid Scintillator Counting) system, equipped with two photomultipliers tubes, was used for CIEMAT/NIST method, using a 3H standard that emits beta particles with maximum energy of 18.7 keV, as efficiency tracing. HIDEX 300SL LSC system, equipped with three photomultipliers tubes, was used for TDCR method. Samples of 14C and 3H, for the liquid scintillator system, were prepared using three commercial scintillation cocktails, UltimaGold, Optiphase Hisafe3 and InstaGel-Plus, in order to compare the performance in the measurements. All samples were prepared with 15 mL scintillators, in glass vials with low potassium concentration. Known aliquots of radioactive solution were dropped onto the cocktail scintillators. In order to obtain the quenching parameter curve, a nitro methane carrier solution and 1 mL of distilled water were used. For measurements in the 4πβ-γ system, 60Co was used as beta gamma emitter. SCS (software coincidence system) was applied and the beta efficiency was changed by using electronic discrimination. The behavior of the extrapolation curve was predicted with code ESQUEMA, using Monte Carlo technique. The 14C activity obtained by the three methods applied in this work was compared and the results showed to be in agreement, within the experimental uncertainty.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2017-03-09
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.