Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.85.2010.tde-08082011-104737
Document
Auteur
Nom complet
Elaine Wirney Martins
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2010
Directeur
Jury
Potiens, Maria da Penha Albuquerque (Président)
Costa, Alessandro Martins da
Rodrigues, Letícia Lucente Campos
Titre en portugais
Estudo e determinação de fatores de influência das dimensões dos frascos de radiofármacos utilizados no IPEN para calibração de ativímetros
Mots-clés en portugais
ativímetros
calibração
controle de qualidade
medicina nuclear
Resumé en portugais
A eficiência e a segurança de uma prática de medicina nuclear dependem, entre outros fatores, de um programa de controle de qualidade, principalmente no que diz respeito à utilização de medidores de atividade de radionuclídeos (ativímetros). Dentre as maiores fontes de erro nas medidas realizadas com um ativímetro estão: a espessura, o tamanho e o volume do frasco que contém o radiofármaco, considerando que um ativímetro típico tem a sua resposta reconhecidamente dependente do frasco utilizado. O propósito desse trabalho foi implantar um programa de controle de qualidade além de estabelecer fatores de correção referente à geometria do frasco de vidro para distribuição de radiofármacos e calibração de ativímetros, considerando que o Laboratório de Calibração de Instrumentos (LCI) do Instituto de Pesquisas Energéticas e Nucleares (IPEN) possui um Sistema Padrão Secundário NPL-CRC radionuclide calibrator, fabricado pela Southern Scientific, composto por uma câmara de ionização tipo poço e um sistema de medida de corrente, com rastreabilidade ao National Physical Laboratory (NPL) e calibrado com um frasco de vidro tipo P6 que possui dimensões diferentes dos utilizados pelo IPEN. Foram testados os radiofármacos 67Ga, 131I, 201Tl e 99mTc, todos produzidos pelo IPEN. Os resultados demonstraram uma variação de até 22% para o radiofármaco 201Tl, sendo que a menor variação foi encontrada para o 131I (2,98%). Os fatores de correção encontrados devem ser incorporados na calibração rotineira dos ativímetros.
Titre en anglais
Study and determination of the influence factors of the radiopharmaceutical vials dimensions used for activimeter calibration at IPEN
Mots-clés en anglais
activimeter
calibration
nuclear medicine
quality control
Resumé en anglais
The efficiency and safety of the nuclear medicine practice depend, among others factors, of a quality control programme, mainly related to the use of the nuclide activity meters (activimeter). One of the most important sources of errors in the activimeter measurements is the thickness, size and volume of the vial that contains the radiopharmaceutical considering that a typical activimeter has its response dependent of the vial used. The objective of this work was to establish a quality control programme and the correction factors for the geometry of the vials used for distribution of radiopharmaceutical and activimeters calibration, considering that the Calibration Laboratory of Instrumentos (LCI) of the Instituto de Pesquisas Energéticas e Nucleares (IPEN) has a NPL-CRC Secondary Standard Radionuclide Calibrator System, manufactured for the Southern Scientific plc, compound by an ionization chamber well type and a current measurement system, with traceability to National Physical Laboratory (NPL) and calibrated with a P6 vial type with different dimensions of the one used for the IPEN. The radiopharmaceutical produced by IPEN 67Ga, 131I, 201Tl and 99mTc, had been tested using the two different vials. The results shown a maximum variation of 22% for 201Tl, and the minimum variation was 2.98% for 131I. The correction factors must be incorporated in the routine calibration of the activimeters.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
2010MartinsEstudo.pdf (1.61 Mbytes)
Date de Publication
2011-08-15
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.