Title in Portuguese
Estudo e determinação de fatores de influência das dimensões dos frascos de radiofármacos utilizados no IPEN para calibração de ativímetros
Keywords in Portuguese
ativímetros
calibração
controle de qualidade
medicina nuclear
Abstract in Portuguese
A eficiência e a segurança de uma prática de medicina nuclear dependem, entre outros fatores, de um programa de controle de qualidade, principalmente no que diz respeito à utilização de medidores de atividade de radionuclídeos (ativímetros). Dentre as maiores fontes de erro nas medidas realizadas com um ativímetro estão: a espessura, o tamanho e o volume do frasco que contém o radiofármaco, considerando que um ativímetro típico tem a sua resposta reconhecidamente dependente do frasco utilizado. O propósito desse trabalho foi implantar um programa de controle de qualidade além de estabelecer fatores de correção referente à geometria do frasco de vidro para distribuição de radiofármacos e calibração de ativímetros, considerando que o Laboratório de Calibração de Instrumentos (LCI) do Instituto de Pesquisas Energéticas e Nucleares (IPEN) possui um Sistema Padrão Secundário NPL-CRC radionuclide calibrator, fabricado pela Southern Scientific, composto por uma câmara de ionização tipo poço e um sistema de medida de corrente, com rastreabilidade ao National Physical Laboratory (NPL) e calibrado com um frasco de vidro tipo P6 que possui dimensões diferentes dos utilizados pelo IPEN. Foram testados os radiofármacos 67Ga, 131I, 201Tl e 99mTc, todos produzidos pelo IPEN. Os resultados demonstraram uma variação de até 22% para o radiofármaco 201Tl, sendo que a menor variação foi encontrada para o 131I (2,98%). Os fatores de correção encontrados devem ser incorporados na calibração rotineira dos ativímetros.
Title in English
Study and determination of the influence factors of the radiopharmaceutical vials dimensions used for activimeter calibration at IPEN
Keywords in English
activimeter
calibration
nuclear medicine
quality control
Abstract in English
The efficiency and safety of the nuclear medicine practice depend, among others factors, of a quality control programme, mainly related to the use of the nuclide activity meters (activimeter). One of the most important sources of errors in the activimeter measurements is the thickness, size and volume of the vial that contains the radiopharmaceutical considering that a typical activimeter has its response dependent of the vial used. The objective of this work was to establish a quality control programme and the correction factors for the geometry of the vials used for distribution of radiopharmaceutical and activimeters calibration, considering that the Calibration Laboratory of Instrumentos (LCI) of the Instituto de Pesquisas Energéticas e Nucleares (IPEN) has a NPL-CRC Secondary Standard Radionuclide Calibrator System, manufactured for the Southern Scientific plc, compound by an ionization chamber well type and a current measurement system, with traceability to National Physical Laboratory (NPL) and calibrated with a P6 vial type with different dimensions of the one used for the IPEN. The radiopharmaceutical produced by IPEN 67Ga, 131I, 201Tl and 99mTc, had been tested using the two different vials. The results shown a maximum variation of 22% for 201Tl, and the minimum variation was 2.98% for 131I. The correction factors must be incorporated in the routine calibration of the activimeters.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
2010MartinsEstudo.pdf (1.61 Mbytes)
Publishing Date
2011-08-15
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.