Desenvolvimento de câmaras de ionização especiais para controle de qualidade em mamografia
Mots-clés en portugais
câmara de ionização
controle de qualidade
dosimetria
mamografia
Resumé en portugais
A mamografia é um método de diagnóstico por imagem que utiliza radiação X. No controle de qualidade dos mamógrafos, as câmaras de ionização são usadas para verificar se os parâmetros de exposição das pacientes estão corretos. Contudo, as câmaras de ionização comerciais para dosimetria em mamografia representam alto custo para clínicas de pequeno e médio porte que desejam ter esse equipamento ou para profissionais que trabalham com controle de qualidade. Assim, a característica inovadora deste trabalho foi desenvolver câmaras de ionização para este fim. Neste trabalho foram projetadas, construídas e caracterizadas câmaras de ionização para feixes de radiação X, no intervalo de energia de mamografia. As câmaras de ionização foram caracterizadas em feixes padronizados de radiação X no LCI/IPEN. Os testes principais de caracterização das câmaras de ionização foram: curva de saturação, linearidade da resposta, dependência energética e angular. Os testes de estabilidade da resposta das câmaras de ionização também foram realizados, apresentando resposta dentro de 2,0 % para estabilidade em longo prazo. Os resultados dos demais testes foram em conformidade com normas internacionais. Essas câmaras de ionização foram ainda submetidas a testes de controle de qualidade de mamógrafos quanto à linearidade das taxas de kerma no ar, à determinação das camadas semirredutoras e das doses glandulares médias. Para a linearidade da taxa de kerma, os valores obtidos foram abaixo dos 10 %, como estabelecido em norma. A dose glandular determinada com as câmaras desenvolvidas apresentaram valores comparáveis com os da câmara de referência testada, com variação prevista em recomendações internacionais.
Titre en anglais
Development of special ionization chambers for a quality control program in mammography
Mots-clés en anglais
dosimetry
ionization chamber
mammography
quality control
Resumé en anglais
Mammography is an imaging method that uses X-rays. The use of ionization chambers in mammography quality control programs presents an essential role which is to verify whether the parameters of the patient exposure are correct. However, the commercial ionization chambers for dosimetry in mammography represent a high cost for small and medium size clinics that wish to have this equipment or for professionals that work with quality control programs. The innovative feature of this work was to develop ionization chambers for this purpose. In this work ionization chambers for X radiation beams in the mammography energy range were designed, constructed and characterized. The ionization chambers were tested in standard X radiation beams at the LCI/IPEN. The main characterization tests performed with the ionization chambers were: saturation curve, linearity of response, angular and energy dependence. The response stability tests of the ionization chambers were also conducted at the LCI, presenting results within 2.0 % for long-term stability. The results of the remaining tests are in accordance with international standards. These ionization chambers were also submitted to quality control tests of mammography equipment: linearity of the air kerma rates, determination of half-value layers and mean glandular doses. The results for air kerma rate linearity were less than 10 %, as recommended in international standards. The mean glandular dose obtained with the developed chambers presented values comparable to those of commercial ionization chambers tested, with an estimated variation within international standards.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2013-07-30
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.