Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.85.2006.tde-03052012-133401
Document
Auteur
Nom complet
Priscilla Roberta Tavares Leite Camargo
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2006
Directeur
Jury
Rodrigues, Laura Natal (Président)
Carvalho, Heloisa de Andrade
Yoriyaz, Hélio
Titre en portugais
Implantação de um programa de controle de qualidade para sistemas de planejamento de tratamento computadorizados de acordo com o TRS 430
Mots-clés en portugais
comissionamento
controle de qualidade
sistemas de planejamento de tratamento computadorizados
Resumé en portugais
No presente trabalho serão apresentadas as diretrizes e os testes necessários para a implantação de um programa de controle de qualidade para o Eclipse 7.3.10 da Varian no Hospital das Clínicas da Faculdade de Medicina da USP, de acordo com a mais recente publicação da AIEA o TRS 430. Os testes recomendados pelo TRS 430 são basicamente divididos em testes de aceitação, comissionamento (testes dosimétricos e não dosimétricos), e testes rotineiros. O documento da AIEA está sendo implementado para o Eclipse no HC para os feixes de fótons de dois aceleradores lineares da Varian, Clinac 600C e Clinac 2100C. Os testes de aceitação verificaram parâmetros de "hardware"; integração do sistema "network"; transferência de dados, e "softwares". Os resultados obtidos mostraram boa concordância com as especificações do fabricante. Para os testes dosimétricos de comissionamento, foram realizadas medidas de dose absoluta para diversos arranjos experimentais. Esses valores foram comparados com os valores de dose gerados pelo SPTC. A grande maioria dos testes apresentou cerca de 90% a 80% dos pontos comparados, dentro dos níveis de tolerância, ou seja, uma boa concordância entre os valores experimentais e os valores gerados pelo SPTC. Somente arranjos de campos assimétricos apresentaram discordâncias grosseiras, mostrando a necessidade de uma investigação mais apurada para esses casos. Os testes de comissionamento não dosimétricos também apresentaram resultados excelentes, com praticamente todas as ferramentas e desempenho geral do sistema de acordo com as recomendações estipuladas no TRS 430. Foram aplicados também critérios de aceitabilidade para a comparação entre os valores de UMs gerados pelo sistema e os valores de UMs calculados manualmente. Os feixes no Eclipse foram caracterizados com dados transferidos do CadPlan e com dados provenientes do recomissionamento dos aceleradores, assim sendo, para esses testes encontrou- se uma diferença de até 3% para campos conformacionados para os dados de feixe provenientes do recomissionamento dos aceleradores, e de até 4% para os dados de feixe transferidos do CadPlan, sendo que o nível de tolerância estabelecido pelo TRS 430 para o arranjo era de 3%.
Titre en anglais
Implementation of a quality assurance program for computerized treatment planning systems according to TRS 430
Mots-clés en anglais
algorithms
dosimetry
experimental data
iaea
linear accelerators
optimization
phantoms
photon beams
planning
quality assurance
quality control
radiation doses
radiotherapy
recommendations
Resumé en anglais
This work presents the guidelines and necessary tests tom implement a quality assurance program for Eclipse 7.3.10 from Varian at Hospital das Clinicas, Sao Paulo University School of Medicine - Brazil, in accordance with the new IAEA publication TRS 430. The recommended tests for the TRS 430 air mainly classified into acceptance tests, commissioning (dosimetric and non-dosimetric tests), and routine tests. The IAEA document's recommendations are being implemented at the hospital for two Varian linear accelerators - Clinac 600C e Clinac 2100C. The acceptance tests verified 'hardware', integration of network systems, data transfer and 'software' parameters. The results obtained are in a good agreement with the manufacturer's specifications. Measurements of absolute dose in several set-ups were made for the commissioning dosimetric tests. These data were compared to the absolute doses determined by the TPS. The great majority of the tests showed 90% to 80% of the analyzed data in acceptance levels, with a good agreement between the experimental data and the data determined by the TPS. Only settings with asymmetric fields presented significant discords, showing the need for a more detailed inquiry for these settings. The non-dosimetric commissioning tests have also presented excellent results, with virtually all the system tools and general performance in compliance with TRS 430. The acceptance criteria have been applied for a comparison between the values of MUs generated by TPS and the calculated manually ones. The beams have been characterized for Eclipse with data transferred from CadPlan and with data from recommissioning of accelerators, so for these tests it was found a difference of at least 3% for the conformal field shape for the data originated in the beams of recommissioning and at least 4% for the data proceeded from CadPlan. The tolerance level established by TRS 430 for this setting was 3%.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2012-05-07
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.