Implementação da irradiação de corpo inteiro em radioterapia
Mots-clés en portugais
comissionamento
dosimetria in vivo
irradiação de corpo inteiro
TLD
Resumé en portugais
Ao implementar uma técnica de tratamento, as características do feixe nas condições de irradiação precisam ser bem conhecidas e estudadas. Cada um dos parâmetros utilizados para cálculo de dose deve ser medido e validado antes da utilização dos mesmos na rotina clínica. Isso se torna ainda mais necessário quando se tratam de técnicas especiais. Neste trabalho estão descritos todos os parâmetros e medidas necessárias para a implementação da irradiação de corpo inteiro numa instalação projetada para tratamentos convencionais, e que faz uso de geometrias não convencionais para gerar os tamanhos de campos alargados. Além disso, são apresentados os dados de comissionamento desta modalidade no Hospital das Clínicas de São Paulo utilizando a comparação de três tipos de detectores para medidas de dose de entrada durante tratamentos de irradiação de corpo inteiro.
Titre en anglais
Implementation of total body irradiation in radiotherapy
Mots-clés en anglais
commissioning
total body irradiation
Resumé en anglais
Before implementing a treatment technique, the characteristics of the beam under irradiation conditions must be well acknowledged and studied. Each one of the parameters used to calculate the dose has to be measured and validated before its utilization in clinical practice. This is particularly necessary when dealing with special techniques. In this work, all necessary parameters and measurements are described for the total body irradiation implementation in facilities designed for conventional treatments that make use of unconventional geometries to generate desired enlarged field sizes. Furthermore, this work presents commissioning data of this modality at Hospital das Clínicas of São Paulo using comparison of three detectors types for measurements of entrance dose during total body irradiation treatment.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2011-08-04
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.