• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.85.2016.tde-01072016-132544
Document
Auteur
Nom complet
Daniele Martins Sarmento
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2016
Directeur
Jury
Carneiro, Janete Cristina Gonçalves Gaburo (Président)
Pinto, Teresa Cristina Nathan Outeiro
Potiens, Maria da Penha Albuquerque
Titre en portugais
Avaliação dos níveis de radiação ambiental no laboratório de tomografia por emissão de pósitrons acoplada a tomografia computadorizada, microPET/CT
Mots-clés en portugais
18F-FDG
controle radiológico
dosimetria termoluminescente
radioproteção
tomografia por emissão de pósitrons (PET/CT)
Resumé en portugais
O sistema microPET/CT é um importante equipamento utilizado nas pesquisas de imagem diagnóstica em pequenos animais. O radiofármaco mais usado nesta tecnologia é o fluordeoxiglicose marcado com flúor-18. Este estudo tem como objetivo efetuar o controle radiológico no laboratório de pesquisa microPET/CT do Centro de Radiofarmácia do IPEN-CNEN/SP, de forma a satisfazer tanto as normas nacionais como as recomendações internacionais. O laboratório está classificado pela equipe de radioproteção da instalação como área supervisionada, nas quais embora não seja obrigatória a adoção de medidas específicas de proteção e segurança, devem ser submetidas reavaliações regulares das condições do ambiente de trabalho. Visando assegurar a proteção radiológica dos trabalhadores diretamente envolvidos no manuseio do equipamento, realizou-se o monitoramento do local de trabalho e a avaliação do controle de dose individual. Inicialmente foi feito o monitoramento pré-operacional, isto é, o levantamento radiométrico no laboratório. Além disso, mediu-se nível de radiação externa nas instalações do laboratório e suas adjacências, por meio da colocação de nove dosímetros termoluminescentes (TL) de CaSO4:Dy, em locais previamente selecionados. Os indivíduos ocupacionalmente expostos foram avaliados mensalmente por meio do uso de dosímetros TL posicionados no tórax e por medidas de corpo inteiro, tomadas a cada seis meses. O período do estudo foi de dois anos, com início em abril de 2014. Para o controle do microPET/CT realizou-se testes de desempenho de acordo com o protocolo padrão do equipamento e em conformidade com a norma desenvolvida pela força tarefa para estudos com PET em animais Animal PET Standard Task Force. O presente estudo permitiu demonstrar que os níveis de radiação das áreas (estimativas de dose ambiente e dose efetiva), assim como a blindagem do equipamento estão adequados de acordo com os limites da exposição ocupacional. Ressalta-se a importância de se seguir rigorosamente os princípios de radioproteção, já que se trata de pesquisas com fontes radioativas não seladas.
Titre en anglais
Evaluation of ambient radiation levels in positron emission tomography/computed tomography in microPET/CT laboratory
Mots-clés en anglais
18F-FDG
positron emission tomography (PET/CT)
radiological control
radioprotection
thermoluminescence dosimetry
Resumé en anglais
Micro PET/CT scanner is an essential tool generally used for small animal molecular imaging. Fluorine-18-labeled fluorodeoxyglucose is the most widely used radioisotope in this technique. The present study aimed to evaluate the radiation levels in a micro PET/CT research laboratory of the Radiopharmacy Center at IPEN-CNEN / SP, in order to accomplish both national standards and international recommendations. The radioprotection team has classified the laboratory as supervised area; even this laboratory does not require the adoption of specific measures for protection and safety, should be done regular re-evaluation of the conditions of occupational exposures. Workplace monitoring and individual control assessment were carried out to ensure the radiological protection of all workers directly involved in handling the scanner. Initially, there was conducted a radiometric survey, as well as measurements of the external radiation level in the workplace and its surroundings. To achieve this goal, there were placed nine thermoluminescent dosimeters of CaSO4:Dy in preselected locations. Monthly evaluations of the occupationally exposed individuals were carried out through the use of TL dosimeters, ported in the workers´ chest. Moreover, whole body measurements were performed every six months. The study period was about two-years which started in April 2014. All tests to evaluate micro PET/CT performance were based on the standard protocol of the equipment in accordance with the standard developed by the Animal PET Standard Task Force. Present study's results demonstrated that the ambient radiation levels (ambient and effective estimated radiation dose), as well as the effective shielding equipment are both adequate. This study emphasizes that it is essential to strictly follow the principles of radioprotection in workplace, whenever researches involve radioactive unsealed sources.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2016-07-01
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.