Disertación de Maestría
DOI
https://doi.org/10.11606/D.85.2010.tde-01062011-122808
Documento
Autor
Nombre completo
Daniel Menezes Dias
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2010
Director
Tribunal
Potiens, Maria da Penha Albuquerque (Presidente)
Nersissian, Denise Yanikian
Vivolo, Vitor
Título en portugués
Estabelecimento de um novo método de calibração de câmaras de ionização tipo lápis para dosimetria em feixes de tomografia computadorizada
Palabras clave en portugués
câmara de ionização tipo lápis
dosimetria em tomografia computadorizada
metrologia em CT
radiodiagnóstico
Resumen en portugués
As câmaras de ionização tipo lápis são utilizadas para dosimetria de feixes em equipamentos de tomografia computadorizada (CT). Neste estudo foi estabelecida uma nova metodologia de calibração destas câmaras, visando a adequação do Laboratório de Calibração de Instrumentos do Instituto de Pesquisas Energéticas e Nucleares (LCI) aos atuais padrões metrológicos internacionais que tratam dos procedimentos específicos para calibração das câmaras utilizadas em CT. Primeiramente partiu-se da implantação das novas qualidades de radiação padrão RQT, que simulam feixes não atenuados usados para aplicações em CT, de acordo com a IEC 61267 da International Electrotechnical Commission (IEC). Após o estabelecimento dos campos padrões de radiação, foi implantada a metodologia de calibração específica para câmaras lápis segundo o Technical Report Series No. 457, da International Atomic Energy Agency (IAEA), que descreve as particularidades do procedimento a ser seguido pelos Laboratórios de Dosimetria Padrão Secundário (SSDLs) em termos de colimação e posicionamento em relação ao feixe de radiação. Inicialmente, foram realizadas medições de PPV (kV) e determinação das filtrações adicionais de cobre por meio dos valores de camada semi-redutora (CSR) recomendados pela norma IEC 61267, para que as qualidades de radiação de referência: RQT 8, RQT 9 e RQT 10 fossem implantadas. Para os filtros adicionais, foram utilizados alumínio e cobre de alta pureza (maior ou igual a 99,9%). Uma metodologia alternativa para determinar as filtrações adicionais também foi utilizada, através da qual se encontrou diretamente as RQTs em espessuras de filtros de cobre equivalente ao conjunto RQR (Al) + Filtração Adicional (Cu), sendo esta uma opção quando não se têm implantadas as RQRs. Com a implantação desta nova metodologia de calibração das câmaras de ionização tipo lápis, o LCI está pronto para calibrar estes instrumentos de acordo com as recentes recomendações internacionais. Permitindo assim uma melhoria na rastreabilidade das calibrações, como também na qualidade dos serviços metrológicos que o IPEN oferece para todo o País.
Título en inglés
Establishment of a new calibration method of pencil ionization chamber for dosimetry in computed tomography
Palabras clave en inglés
dosimetry in computed tomography
pencil ionization chamber
Resumen en inglés
Pencil ionization chambers are used for beam dosimetry in computed tomography equipment (CT). In this study, a new calibration methodology was established, in order to make the Calibration Laboratory of Instituto de Pesquisas Energéticas e Nucleares (LCI) suitable to international metrological standards, dealing with specific procedures for calibration of these chambers used in CT. Firstly, the setup for the new RQT radiation qualities was mounted, in agreement with IEC61267 from the International Electrotechnical Commission (IEC). After the establishment of these radiation qualities, a specific calibration methodology for pencil ionization chambers was set, according to Technical Report Series No. 457, from the International Atomic Energy Agency (IAEA), which describes particularities of the procedure to be followed by the Secondary Standard Dosimetry Laboratories (SSDLs), concerning to collimation and positioning related to the radiation beam. Initially, PPV (kV) measurements and the determination of copper additional filtrations were carried out, measuring the half value layers (HVL) recommended by the IEC 61267 standard, after that the RQT 8, RQT 9 and RQT 10 radiation quality references were established. For additional filters, aluminum and copper of high purity (around 99.9%) were used. RQTs in thickness of copper filters equivalent to the set RQR (Al) + Additional Filtration (Cu) was directly found by an alternative methodology used to determine additional filtrations, which is a good option when RQRs have not the possibility of be setting up. With the establishment of this new methodology for the ionization pencil chambers calibration, the LCI is ready to calibrate these instruments according to the most recent international standards. Therefore, an improvement in calibration traceability, as well as in metrological services offered by IPEN to all Brazil is achieved.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2011-06-21
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.