Doctoral Thesis
DOI
https://doi.org/10.11606/T.82.2013.tde-16062014-153409
Document
Author
Full name
José Ricardo Camilo Pinto
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Carlos, 2013
Supervisor
Committee
Rollo, João Manuel Domingos de Almeida (President)
Fortulan, Carlos Alberto
Morais, Dionei Freitas de
Rigo, Eliana Cristina da Silva
Vieira, Edson Del Rio
Title in Portuguese
Avaliação hidrodinâmica de uma válvula neurológica ajustável por acionamento mecânico
Keywords in Portuguese
Biomateriais
Derivação ventrículo-peritoneal
Hidrocefalia
Líquor
Teste hidrodinâmico
Válvula neurológica
Abstract in Portuguese
A hidrocefalia é uma doença ocasionada pelo distúrbio na formação, circulação ou absorção do líquido cefalorraquidiano (líquor) que acarreta na elevação da pressão intracraniana. O tratamento mais usual para a disfunção é a derivação ventrículo-peritoneal (DVP) responsável pela drenagem do líquor do ventrículo até o abdômen do paciente, através do implante de um cateter ventricular, uma válvula neurológica e um cateter peritoneal. Desse modo, o presente estudo apresenta o conceito de uma nova válvula neurológica com um mecanismo inovador para alteração da pressão de funcionamento, com um menor custo produtivo e de possível aceitação pelo Sistema Único de Saúde (SUS). A norma ISO 7197 forneceu as diretrizes atendidas no desenvolvimento da válvula neurológica ajustável e as informações para a construção das bancadas de testes, utilizadas para a realização dos ensaios hidrodinâmicos no protótipo do dispositivo valvular. Foram pesquisados os biomateriais empregados na fabricação de válvulas neurológicas comerciais, para a proposição dos materiais necessários para a construção da válvula ajustável por acionamento mecânico proposta. O aparato utilizado nos testes hidrodinâmicos apresentou incertezas experimentais, que não comprometem os resultados e a avaliação do desempenho in vitro do dispositivo testado. O protótipo da válvula resistiu aos testes de vazamento e de refluxo de fluido. As pressões de abertura e de fechamento apresentaram-se adequadas para o início e a interrupção do escoamento na DVP. Também foi obtido o comportamento do diferencial de pressão, em função da vazão, para as quatro classes de funcionamento disponíveis no protótipo testado. A avaliação hidrodinâmica do protótipo da válvula ajustável por acionamento mecânico apresenta resultados satisfatórios no controle da drenagem de líquor, frente a todos os ensaios hidrodinâmicos realizados.
Title in English
Hydrodynamic evaluation of an adjustable neurological valve by mechanical drive
Keywords in English
Biomaterials
CSF
Hydrocephalus
Hydrodynamic test
Neurological valve
Shunt
Abstract in English
Hydrocephalus is a disease caused by disturbance in formation, circulation or absorption of cerebrospinal fluid (CSF) which causes high intracranial pressure. The most common treatment for the dysfunction is the ventricle-peritoneal shunt responsible for drainage of cerebrospinal fluid from the patient's ventricle to the abdomen, through the implant of a ventricular catheter, a neurological valve and a peritoneal catheter. Thus, the present study introduces the concept of a new valve with an innovative mechanism to change the working pressure, with a lower cost of production and possible acceptance by the Brazilian Health System. ISO 7197 provided the guidelines met in the development of adjustable neurological valve and the information for the construction of testing rigs, used for hydrodynamic testing on the valve device prototype. Biomaterials used in manufacturing commercial neurological valves have been researched for proposition of materials needed for the proposed adjustable neurological valve by mechanical drive. The apparatus used in hydrodynamic tests presented experimental uncertainties, which do not compromise the results and evaluation of in vitro performance of the device tested. The prototype of the valve withstood the leak and the fluid backflow testing. The opening and closing pressures were suitable for the beginning and the interruption of the flow in the ventricle-peritoneal shunt. It was also obtained the behavior of the pressure differential, due to the flow, for the four classes of operation available in the prototype tested. Hydrodynamic evaluation of the prototype of the adjustable valve by mechanical drive shows satisfactory results in the control of CSF drain, dealing with all hydrodynamic tests carried out.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2014-06-18
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.