Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.82.2008.tde-15122008-101842
Document
Auteur
Nom complet
Nicola Bloise
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Carlos, 2008
Directeur
Jury
Silva, Orivaldo Lopes da (Président)
Castro Neto, Jarbas Caiado de
Marques, Paulo Mazzoncini de Azevedo
Titre en portugais
Investigação da possibilidade de correlação dos potenciais transcutâneos visuais evocados com o alcance do campo da vista humana
Mots-clés en portugais
Campimetria
Campo visual
PVE
Visão
Resumé en portugais
Este trabalho descreve o desenvolvimento de um instrumento automático para medida de campo visual do olho humano, baseado na ocorrência de potenciais visuais evocados O principal objetivo do estudo foi demonstrar que os potenciais visuais evocados (P100) observados na região occipital podem ser correlacionados à visão de flashes de fontes luminosas pontuais distribuídas sobre uma calota com a forma de um elipsóide deformado. Aplicou-se ao domo hemisférico uma distribuição de fontes luminosas pontuais que descrevem a forma do campo visual humano (limitado a 60º acima, 70º abaixo do eixo central, limitado a 60º no lado nasal e 105º no lado temporal de cada olho). A ausência de potenciais visuais evocados por flashes luminosos é utilizada como indicador de perda da sensitividade da retina. Detalhes do desenvolvimento deste instrumento e os resultados preliminares são apresentados.
Titre en anglais
Inquiry about the possibility of correlation among the visual transcutâneos evoked visual potentials and the reach of the human visual field
Mots-clés en anglais
Campimetry
EVP
Vision
Visual field
Resumé en anglais
An instrument for automatic in vivo evaluation of the human visual field boundaries based on the evoked visual potentials was developed. The main objective of this study is to demonstrate that evoked potentials (P100) over the occipital area can be correlated to the vision of punctual light flashes from a deformed ellipsoid-shaped calotte. A punctual light sources distribution that follows the human visual field (boundaries lying 60º upward, 70º downward from the central axis, limited to 60º in the nasal side and 105º on the temple side of each eye) was applied to the hemispherical dome. The absences of light flashes evoked potentials are used as an indicator of lack the sensivity of the retina. The details of the development of this instrument and preliminary results for in vivo eyes are presented.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
TDE_NicolaBloise.pdf (3.19 Mbytes)
Date de Publication
2010-04-28
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.