Tesis Doctoral
DOI
https://doi.org/10.11606/T.76.2010.tde-06122010-141123
Documento
Autor
Nombre completo
Gustavo Henrique Frigieri Vilela
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Carlos, 2010
Director
Tribunal
Oliveira, Sergio Mascarenhas (Presidente)
Machado, Benedito Honório
Sameshima, Koichi
Tannus, Alberto
Verzola, Roberto Mario Machado
Título en portugués
Desenvolvimento de um sistema minimamente invasivo para monitorar a pressão intracraniana
Palabras clave en portugués
Extensômetro
Instrumentação médica
Monitoramento minimamente invasivo
PIC
Pressão Intracraniana
Resumen en portugués
A pressão intracraniana (PIC) é um dos principais parâmetros neurológicos em animais e humanos. A PIC é uma função da relação entre o conteúdo da caixa craniana (parênquima cerebral, líquido cefalorraquiano e sangue) e o volume do crânio. O aumento da PIC (hipertensão intracraniana) pode acarretar graves efeitos fisiológicos ou até mesmo o óbito em pacientes que não receberem rapidamente os devidos cuidados, os quais incluem o monitoramento em tempo real da PIC. Todos os métodos de monitoramento da PIC atualmente utilizados são invasivos, ou seja, é necessário introduzir um sensor de pressão no sistema nervoso central, acarretando aos pacientes riscos de infecções e traumas decorrentes do método. Neste trabalho desenvolvemos um método minimamente invasivo de monitoramento da pressão intracraniana, que consiste na utilização de sensores de deformação do tipo strain gauge fixados sobre a calota craniana. Os sinais oriundos deste sensor foram amplificados, filtrados e enviados para um computador com software apropriado para análise e armazenamento dos dados. O trabalho aqui apresentado objetivou os testes "in vivo" do sistema, onde foi utilizada mais de uma centena de animais em diversos testes, sendo que em todos os casos os resultados foram satisfatórios, apontando a eficácia do método.
Título en inglés
Development of a minimally invasive system to monitor the intracranial pressure
Palabras clave en inglés
ICP
Intracranial Pressure
Medical instrumentation
Minimally invasive monitoring
Strain gauge
Resumen en inglés
The intracranial pressure (ICP) is one of the most important neurological parameter in animals and humans. The ICP is a function of the relation between the contents of the skull (brain parenchyma, cerebrospinal fluid and blood) and the volume of the skull. The increase in ICP (intracranial hypertension) may cause serious physiological effects and death in patients that do not receive appropriate care quickly, which includes real-time monitoring of ICP. All monitoring methods currently used in ICP are invasive, ie requiring invasion of the central nervous system by a pressure sensor, causing infections and traumas risks to patients. In this work we present a new minimally invasive method to monitor the intracranial pressure. This uses strain gauge deformation sensors, externally glued on the skull. The signal from this sensor is amplified, filtered and sent to a computer with appropriate software for analysis and data storage. "In vitro" and "in vivo" experiments let to the following results: (1) Our minimally invasive system is capable of adequately monitoring the ICP. (2) The measurements are in real and online time providing excellent signal and stability. (3) Simultaneous comparison with invasive methods not only validated our results but showed increased performance. The equipment cost effective will allow the use of our system in the Public Health System, with a important social aspect of our contribution.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2010-12-07
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.