• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.75.2006.tde-05062007-142325
Document
Auteur
Nom complet
Flavio de Oliveira Leme
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Carlos, 2006
Directeur
Jury
Krug, Francisco Jose (Président)
Mazo, Luiz Henrique
Nobrega, Joaquim de Araujo
Titre en portugais
"Avaliação do método das médias móveis no tratamento de sinais de absorção atômica de cádmio, chumbo e crômio obtidos por atomização eletrotérmica em filamento de tungstênio"
Mots-clés en portugais
cádmio
chumbo
crômio
filamento de tungstênio
médias móveis
Resumé en portugais
Este trabalho teve como objetivo avaliar o uso do método das médias móveis na atomização de cádmio, crômio e chumbo utilizando filamento de tungstênio de 150 W (OSRAM) como atomizador eletrotérmico. Realizou-se um estudo sistemático da influência da altura de observação, da temperatura e taxa de aquecimento na atomização dos analitos e a viabilidade do uso do método das médias móveis no tratamento dos sinais transientes gerados na atomização. O estudo da influência da altura de observação foi realizado posicionado-se o filamento no centro (0 mm), 1 e 2 mm abaixo do feixe de radiação. A influência da temperatura e da taxa de aquecimento foi estudada utilizando temperaturas na etapa de atomização entre 1890 e 2780 °C. O método das médias móveis foi aplicado aos sinais transientes gerados na atomização, com o filamento posicionado no centro (0 mm), 1 e 2 mm e com temperatura de atomização entre 2060 e 2780 °C, utilizando fatores 0, 3, 5, 7 e 9. No estudo da temperatura e da taxa de aquecimento foram aplicadas ddps no software de controle na etapa de atomização entre 4 e 12 V. As soluções usadas no desenvolvimento deste trabalho foram preparadas em meio 0,1 % v/v HNO3. Utilizou-se como gás de proteção a mistura de 90% Ar+10 % H2. Para o cádmio, recomenda-se a utilização da altura de observação de 1 mm, para o chumbo recomenda-se 0 mm e para o crômio a altura recomendada é a de 2 mm. Os estudos comprovaram a dependência dos sinais de absorção atômica em altura de pico com a temperatura e a taxa de aquecimento. O cádmio foi o analito em que a taxa de aquecimento foi a menos crítica para a atomização, enquanto que o crômio foi o mais afetado pela variação da taxa de aquecimento. Não foi observada diferença significativa entre os sinais de cádmio com alturas de observação de 0 e 1 mm. Para o chumbo, a diferença entre os sinais com altura de observação de 0 e 1 mm foi constante em todas as taxas de aquecimento estudadas. A 2 mm a diferença entre os sinais aumentou com o aumento da taxa de aquecimento. Para o crômio, não ocorreu uma diferença significativa entre os sinais nas alturas de 1 e 2 mm com atomização até 2180 °C. Nos três analitos a sensibilidade aumentou com o aumento da taxa de aquecimento. O método das médias móveis permitiu a diminuição da variação dos sinais transientes do branco e do ruído. Para o cádmio, não ocorreu uma melhora no limite de detecção do método com a aplicação das médias móveis em nenhuma condição. Para o chumbo e crômio, ocorreu uma melhora no limite de detecção do método nas alturas de observação de 0 e 1 mm. O limite de detecção instrumental apresentou uma melhora significativa para os três elementos em todas as condições estudadas.
Titre en anglais
"Evaluation of the moving averages in the treatment of cadmium, lead and chromium atomic absorption signals from electrothermal atomization"
Mots-clés en anglais
cadmium
chromium
electrothermal atomizer
lead
moving averages
Resumé en anglais
The aim of this paper was to evaluate the use of moving averages of cadmium, chromium and lead transient atomic absorption signals generated by 150 W (OSRAM) tungsten coil electrothermal atomizer. A systematic study dealing with the influence of observation height, temperature and heating rate on atomization was carried through in order to demonstrate the viability of moving averages in the treatment of the transient signals during the atomization step. The influence of observation height was carried out with the tungsten coil in the center (0 mm), at 1 and 2 mm below the radiation beam. The influence of atomization temperature and heating rate were evaluated between 1890 °C (3.3 K.ms-1) and 2780 °C (9,3 K.ms-1). The transient moving averages were obtained by using factors 0, 3, 5, 7 and 9, with the tungsten coil in the center (0 mm), 1 and 2 mm and with atomization between 2060 and 2780 °C. For obtaining different temperatures and heating rates, an appropriate power supply was used by applying 4 to 12 V (40 to 120 W) to the atomizer during atomization and a mixture 90% Ar+10% H2 was used as purge/atomization gas. Blanks and reference solutions were prepared in 0,1% v/v HNO3. Best results for cadmium were obtained at 1 mm. For lead it is recommended 0 mm and for chromium 2 mm. The studies had proven the dependence of peak height absorbance signals with temperature and heating rate. Cadmium was the analyte where the heating rate was to less critical for the atomization, whereas chromium was affected by the variation of the heating rate. Significant differences were not observed between cadmium signals with observation heights of 0 and 1 mm. For lead, the difference between the absorbance signals with observation heights of 0 and 1 mm were constant in all tested heating rates. At 2 mm the difference between signals increased by increasing the heating rates. For chromium, a significant difference was not noticed between peak height absorbance signals at 1 and 2 mm with up to 2180 °C. As expected, sensitivity increased for all three analytes by increasing the heating rate. The moving average is a powerful mathematical treatment for minimizing unpredictable fluctuations of the base line during the chosen time interval for measuring the transient absorbance signals. With the exception of Cd, where no improvements on the method detection limit were noticed, the application of the moving average for the lead and chromium showed 3 to 4 fold improvement in the LDmet at 0 and 1 mm observation heights, respectively. The LDinstr presented a significant improvement (3 to 6 fold) for the three analytes in all studied conditions.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2007-06-11
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.