Master's Dissertation
DOI
https://doi.org/10.11606/D.64.2011.tde-30092011-100908
Document
Author
Full name
Jeová Correia Miranda
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
Piracicaba, 2011
Supervisor
Committee
Reis, Boaventura Freire dos (President)
Masini, Jorge Cesar
Teixeira, Leonardo Sena Gomes
Title in Portuguese
Desenvolvimento de um equipamento portátil e de sistema de análises em fluxo empregando multicomutação. Determinação fotométrica de ferro em águas de rios
Keywords in Portuguese
Águas naturais
Análises em fluxo
Espectrofotomeria
Ferro
Fotômetro de LED
Mini-bombas solenóide
Multicomutação
Abstract in Portuguese
O ferro é o elemento mais abundante na Terra e tem importante papel em ciclos biogeoquímicos Neste trabalho foi desenvolvido um procedimento analítico automático para a determinação fotométrica de ferro em águas de rios, empregando um fotômetro baseado em LED e mini-bombas solenoide para o gerenciamento de soluções. O procedimento proposto foi baseado na reação do ferro (III) com o tiocianato de potássio formando um complexo avermelhado com máximo de absorção em 470 nm. Para a determinação de ferro total foi incluída uma etapa de oxidação em linha, usando uma solução de persulfato (S2O82-) como oxidante. O módulo de análises foi constituído por quatro mini-bombas solenoide e uma válvula solenoide de três vias. Para a obtenção de sinais foi construído um fotômetro composto por uma cela de fluxo com caminho óptico de 50 mm, um LED azul ('lâmbda'=470 nm) e um fotodetector. Visando utilizar este equipamento portátil para medidas em campo, foram empregadas duas baterias de 6 V para alimentar os dispositivos ativos. Os ensaios em laboratório mostraram que as baterias permitem uma autonomia de mais de 50 horas de trabalho sem recarregá-las. O procedimento desenvolvido apresentou as seguintes características analíticas: resposta linear entre as concentrações de 0,25 e 4,0 mg L-1, segundo a equação Y = (0,0183 ± 0,003) + (0,15326 ± 0,002)X (r=0,9997); limite de detecção 0,013 mg L-1 Fe (III) estimado com 99,7 % de confiança; desvio padrão relativo estimado em 0,95 % (n=15) obtido com uma solução de concentração de 1,5 mg L-1 Fe (III); frequência de amostragem de 50 determinações por hora. Os resultados obtidos pelo procedimento proposto e pelo de referência não apresentaram diferenças significativas com nível de confiança de 95 %
Title in English
Development of a portable equipment and a flow analysis system employing multicommutation. Photometric determination of iron in river waters
Keywords in English
Flow analysis
Iron
LED photometer
Multicommutation
Natural waters
Solenoid mini-pumps
Spectrophotometry
Abstract in English
Iron is the most abundant element on Earth and plays an important role in biogeochemical cycles. In this work, an automatic analytical procedure was developed for the photometric determination of iron in river waters, employing a LED-based photometer and solenoid mini-pumps for solution handling. The proposed procedure was based on the reaction between iron (III) and potassium thiocyanate, yielding a red complex with maximum absorption at 470 nm. For the determination of total iron, an in line oxidation step was included, employing a persulfate solution (S2O82-) as oxidant. The module was constructed with four solenoid mini-pumps and a three-way solenoid valve. For data acquisition a photometer was constructed with a 50 mm optical path flow cell, a blue LED ('lâmbda'= 470 nm) and a photodetector. Envisioning the use of this portable equipment in field, two 6-Volt batteries were employed to supply energy to the active devices. Laboratory tests showed that the batteries provide more than 50 hours of operation without recharging. The proposed procedure showed the following analytical features: linear response between 0.25 and 4.0 mg L-1, according to the equation Y = (0.0183 ± 0.003) + (0.15326 ± 0.002) X (r = 0.9997); detection limit of 0.013 mg L-1 Fe (III), estimated with 99.7 % of confidence; relative standard deviation estimated as 0.95% (n = 15), obtained with a 1.5 mg L-1 Fe (III) solution; sampling rate of 50 determinations per hour.The results obtained with the proposed and reference procedure did not show significative differences at a 95% confidence level
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.