Doctoral Thesis
DOI
https://doi.org/10.11606/T.64.2010.tde-10092010-160704
Document
Author
Full name
Marcelo Alves Teixeira
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
Piracicaba, 2010
Supervisor
Committee
Reis, Boaventura Freire dos (President)
Cavalheiro, Eder Tadeu Gomes
Faria, Ronaldo Censi
Fatibello Filho, Orlando
Nogueira, Ana Rita de Araujo
Title in Portuguese
Desenvolvimento de instrumentação e procedimentos automáticos para determinação de arsênio e fluoreto em águas empregando multicomutação em fluxo e detecção fotométrica
Keywords in Portuguese
Água
Análise por injeção em fluxo
Arsênio
Fluoreto
Microcontrolador
Minibombas
Multicomutação
Abstract in Portuguese
Neste trabalho, é proposto o desenvolvimento de instrumentação e procedimentos analíticos automáticos para a determinação de fluoreto e arsênio em águas empregando o conceito de multicomutação em fluxo e detecção fotométrica. Na determinação de fluoreto, o controle dos dispositivos foi realizado empregando um computador, programado em QuickBasic 4.5. O fotômetro desenvolvido utilizava um fotodetector (OPT301), uma cela de fluxo com 60 mm de caminho óptico e um LED ('lambda' = 525 nm), usado como fonte de radiação. Os sinais gerados pelo fotômetro foram convertidos para digital empregando um multímetro digital, que os transferia para o computador através da interface serial RS-232. O procedimento para a determinação de fluoreto foi baseado no método de zircônio/ vermelho de alizarina S, onde o fluoreto retira o zircônio da estrutura metal ligante formando um complexo incolor. Após a seleção das variáveis de controle, o procedimento proposto apresentou resposta linear na faixa de 0,25 a 1,75 mg L-1 (LD = 0,06 mg L-1), desvio relativo menor que 3,5%, freqüência de amostragem de 45 h-1. No sistema para determinação de arsênio em águas, o controle do módulo de análise foi realizado utilizando um microcontrolador da série PIC 18F2553, o qual contém um conversor A/D interno de 12 bits de resolução e capacidade de operar com taxa de amostragem de até 100.000 s-1. O mesmo recebia os parâmetros de controle por um computador via USB 2.0, que utilizava a mesma interface para transferir os dados provenientes do fotômetro. O software para essa comunicação foi escrito em Delphi 7.0 e a linguagem C foi utilizada para a programação do microcontrolador. Como sistema de detecção foi o mesmo empregado na determinação de fluoreto, utilizando um LED ('lambda' = 625 nm) como fonte de radiação. O procedimento proposto para a determinação de arsênio foi baseado na reação do arsênio inorgânico com boro-hidreto, onde a arsina formada reagia com azul de metileno em meio micelar. A descoloração da solução de azul de metileno era proporcional à concentração de arsênio presente na amostra. A procedimento proposto apresentou resposta linear na faixa de 2,5 a 40 'mü'g L-1 (r = -0,995), limite de detecção de 0,3 'mü'g L-1, desvio relativo menor que 6,5 %, freqüência de amostragem de 25 h-1.
Title in English
Development of instrumentation and automatic procedures for the determination of arsenic and fluoride in waters by using flow multicommutaion and photometric detection
Keywords in English
Arsenic
Flow injection analysis
Fluoride
Microcontroller
Mini-pumps
Multicommutation
Water
Abstract in English
The development of instrumentation and automatic procedures for the determination of arsenic and fluoride in waters by using flow multicommutation and photometric detection were proposed. For fluoride determination, the system was controlled by a microcomputer equipped with MS-Dos environment and running a software written in Quick Basic 4.5. It was used a homemade photometric detector OPT301, a 60 mm pathway flow cell, and a light emitting diode ('lambda' = 525 nm) as light source. Data acquisition was performed by using a digital multimeter and a microcomputer with RS- 232 serial interface. The method was based on a complexation reaction between fluoride and alizarin red S, where the fluoride takes out the zircon from the metal-binding moieties resulting in color bleaching. After optimization of parameters, the procedure provided linear response in the 0.25 to 1.75 mg L-1 range, limit of detection of 0.06 mg L- 1, coefficient of variation lower than 3.5 %, and sample throughput of 45 determinations per hour. For arsenic determination, the control module was performed by using a PIC 18F2553 microcontroller equipped with a 12 bits resolving power A/D internal converter, working at 100,000 sampling per second rate. Data acquisition was performed by a microcomputer equipped with Windows environment and running a software written in Delphi 7.0. A LED ('lambda' = 625 nm), as radiation source, was used for As determinations by using the same detection system used for fluoride determination. The method for As determination was based on the color bleaching of methylene blue in micellar media by AsH3 produced from inorganic As and NaBH4. The color bleaching of methylene blue was proportional to As concentration in the sample. The proposed method provided linear response in the 2.5 to 40 'mü'g L-1 range, limit of detection of 0,3 'mü'g L-1, coefficient of variation lower than 6.5 %, and sample throughput of 25 determinations per hour.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Tese_TeixeiraMA.pdf (7.92 Mbytes)
Publishing Date
2011-02-03
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.