Disertación de Maestría
DOI
https://doi.org/10.11606/D.64.2016.tde-02022016-110804
Documento
Autor
Nombre completo
Eliézer de Oliveira
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
Piracicaba, 2015
Director
Tribunal
Zagatto, Elias Ayres Guidetti (Presidente)
Carrilho, Elma Neide Vasconcelos Martins
Kamogawa, Marcos Yassuo
Título en portugués
Determinação espectrofotométrica catalítica de molibdênio e ferro explorando calibração multivariada em um sistema de análises químicas em fluxo
Palabras clave en portugués
Análise cinética diferencial
Análises químicas por injeção em fluxo
Calibração multivariada
Espectrofotometria
Molibdênio e ferro
Resumen en portugués
Os efeitos catalíticos de Mo(VI) e Fe(III) na oxidação de iodeto por peróxido de hidrogênio foram explorados visando à determinação simultânea destes. Diferenças nas taxas de reação foram alcançadas através do estabelecimento de duas zonas reacionais distintas para alteração das cinéticas reacionais. Para isto, foi projetado um sistema de análises por injeção em fluxo que permite a inserção de duas alíquotas de amostra, sendo adicionado ácido sulfúrico em uma destas com formação de uma zona de amostra complexa. Cada elemento de fluido desta zona é caracterizado por composição química, tempo de residência e taxa de reação específicas. Os principais parâmetros que afetam a taxa reacional, tais como concentração dos reagentes, comprimento do reator e vazão total, dentre outros, foram avaliados buscando melhoria na sensibilidade e discriminação cinética entre os analitos. Os comprimentos das alças de amostragem e da bobina separadora foram avaliados construindo-se modelos matemáticos visando máxima discriminação nas previsões para os analitos. Os modelos foram construídos a partir da ferramenta quimiométrica PLS (Partial Least Squares) utilizando os dados gerados por medidas de máximos e mínimos ao longo da zona de amostra. O conjunto de calibração era composto por misturas de Mo(VI) e Fe(III) na faixa de concentrações 6,2 - 50,0 ?g L-1 e 0,5 - 7,0 mg L-1, respectivamente. O modelo selecionado apresenta boa habilidade de previsão com valores de RMSEP estimados como 0,67 ?g L-1 para Mo(VI) e 0,21 mg L-1 para Fe(III). Recuperações entre 85 e 114% foram obtidas para amostras de águas naturais. O sistema proposto apresenta estabilidade física, boa repetibilidade entre as leituras (d. p. r. < 2 %), velocidade analítica de 31 h-1 e consumo de 54 mg de iodeto de potássio e 0,30 mg de peróxido de hidrogênio por amostra
Título en inglés
Catalytic spectrophotometric determination of Mo(VI) and Fe(III) exploiting multivariate calibration implemented in a flow system
Palabras clave en inglés
Differential kinetic analysis
Flow injection analysis
Molybdenum and iron
Multivariate calibration
Spectrophotometry
Resumen en inglés
The catalytic effects of Mo(VI) and Fe(III) in the oxidation of iodide by hydrogen peroxide were exploited aiming at their simultaneous determination. Differences in reaction rates were achieved by establishing of two distinct reactions zones to modify the reaction kinetics involved. To this end, a flow injection system allowing insertion of two sample aliquots was designed, and sulfuric acid was added to one of the aliquots. A complex sample zone was then established. Each fluid element of this zone was characterized by a given chemical composition, residence time and reaction rate. The main parameters affecting the reaction rate, such as reagent concentrations, analytical path length and total flow rate, among others, were evaluated for improving sensitivity and analyte kinetic discrimination. Influences of sampling loop lengths and distance between initial plugs were studied, and mathematical models were constructed aiming at improved discrimination in prediction of analytes. The models were constructed from the PLS (Partial Least Squares) chemometric tool using data generated by measurements performed on sample zone regions with maximum and minimum local concentrations. The calibration set consisted of Mo(VI) and Fe(III) mixtures in concentration ranges of 6.2 - 50.0 ?g L-1 and 0.5 - 7.0 mg L-1, respectively. The selected model is characterized by good prediction ability prediction, as RMSEP values were estimated as 0.67 ?g L-1 for Mo(VI) and 0.21 mg L-1 for Fe(III), respectively. Recoveries within the 85 - 114% range were obtained for natural waters. The proposed system presents physical stability and favorable characteristics of measurement repeatability (r.s.d.<2%), sampling rate (31 h-1) and reagent consumption (54 mg of KI and 0.30 mg H2O2 per sample)
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2016-03-10
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.