Master's Dissertation
DOI
https://doi.org/10.11606/D.64.2016.tde-02022016-110804
Document
Author
Full name
Eliézer de Oliveira
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
Piracicaba, 2015
Supervisor
Committee
Zagatto, Elias Ayres Guidetti (President)
Carrilho, Elma Neide Vasconcelos Martins
Kamogawa, Marcos Yassuo
Title in Portuguese
Determinação espectrofotométrica catalítica de molibdênio e ferro explorando calibração multivariada em um sistema de análises químicas em fluxo
Keywords in Portuguese
Análise cinética diferencial
Análises químicas por injeção em fluxo
Calibração multivariada
Espectrofotometria
Molibdênio e ferro
Abstract in Portuguese
Os efeitos catalíticos de Mo(VI) e Fe(III) na oxidação de iodeto por peróxido de hidrogênio foram explorados visando à determinação simultânea destes. Diferenças nas taxas de reação foram alcançadas através do estabelecimento de duas zonas reacionais distintas para alteração das cinéticas reacionais. Para isto, foi projetado um sistema de análises por injeção em fluxo que permite a inserção de duas alíquotas de amostra, sendo adicionado ácido sulfúrico em uma destas com formação de uma zona de amostra complexa. Cada elemento de fluido desta zona é caracterizado por composição química, tempo de residência e taxa de reação específicas. Os principais parâmetros que afetam a taxa reacional, tais como concentração dos reagentes, comprimento do reator e vazão total, dentre outros, foram avaliados buscando melhoria na sensibilidade e discriminação cinética entre os analitos. Os comprimentos das alças de amostragem e da bobina separadora foram avaliados construindo-se modelos matemáticos visando máxima discriminação nas previsões para os analitos. Os modelos foram construídos a partir da ferramenta quimiométrica PLS (Partial Least Squares) utilizando os dados gerados por medidas de máximos e mínimos ao longo da zona de amostra. O conjunto de calibração era composto por misturas de Mo(VI) e Fe(III) na faixa de concentrações 6,2 - 50,0 ?g L-1 e 0,5 - 7,0 mg L-1, respectivamente. O modelo selecionado apresenta boa habilidade de previsão com valores de RMSEP estimados como 0,67 ?g L-1 para Mo(VI) e 0,21 mg L-1 para Fe(III). Recuperações entre 85 e 114% foram obtidas para amostras de águas naturais. O sistema proposto apresenta estabilidade física, boa repetibilidade entre as leituras (d. p. r. < 2 %), velocidade analítica de 31 h-1 e consumo de 54 mg de iodeto de potássio e 0,30 mg de peróxido de hidrogênio por amostra
Title in English
Catalytic spectrophotometric determination of Mo(VI) and Fe(III) exploiting multivariate calibration implemented in a flow system
Keywords in English
Differential kinetic analysis
Flow injection analysis
Molybdenum and iron
Multivariate calibration
Spectrophotometry
Abstract in English
The catalytic effects of Mo(VI) and Fe(III) in the oxidation of iodide by hydrogen peroxide were exploited aiming at their simultaneous determination. Differences in reaction rates were achieved by establishing of two distinct reactions zones to modify the reaction kinetics involved. To this end, a flow injection system allowing insertion of two sample aliquots was designed, and sulfuric acid was added to one of the aliquots. A complex sample zone was then established. Each fluid element of this zone was characterized by a given chemical composition, residence time and reaction rate. The main parameters affecting the reaction rate, such as reagent concentrations, analytical path length and total flow rate, among others, were evaluated for improving sensitivity and analyte kinetic discrimination. Influences of sampling loop lengths and distance between initial plugs were studied, and mathematical models were constructed aiming at improved discrimination in prediction of analytes. The models were constructed from the PLS (Partial Least Squares) chemometric tool using data generated by measurements performed on sample zone regions with maximum and minimum local concentrations. The calibration set consisted of Mo(VI) and Fe(III) mixtures in concentration ranges of 6.2 - 50.0 ?g L-1 and 0.5 - 7.0 mg L-1, respectively. The selected model is characterized by good prediction ability prediction, as RMSEP values were estimated as 0.67 ?g L-1 for Mo(VI) and 0.21 mg L-1 for Fe(III), respectively. Recoveries within the 85 - 114% range were obtained for natural waters. The proposed system presents physical stability and favorable characteristics of measurement repeatability (r.s.d.<2%), sampling rate (31 h-1) and reagent consumption (54 mg of KI and 0.30 mg H2O2 per sample)
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2016-03-10
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.