• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Thèse de Doctorat
DOI
https://doi.org/10.11606/T.60.2016.tde-06072016-104515
Document
Auteur
Nom complet
Aline Akemi Ishikawa
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
Ribeirão Preto, 2016
Directeur
Jury
Gaitani, Cristiane Masetto de (Président)
Paschoal, Jonas Augusto Rizzato
Santos Neto, Álvaro José dos
Barth, Thiago
Jager, Alessandra Vincenzi
Titre en portugais
Desenvolvimento e validação de método para análise do topiramato em amostras de plasma por eletroforese capilar com detecção UV/vis e C4D
Mots-clés en portugais
C4D
eletroforese capilar
plasma
topiramato
UV/vis.
Resumé en portugais
O topiramato (TPM) é um fármaco antiepiléptico utilizado para o controle de crises epilépticas generalizadas e parciais em adultos e crianças. A determinação da concentração plasmática do TPM em associação com as observações clínicas é de suma importância, uma vez que permite o ajuste individual da dose administrada ao paciente, diminuindo a incidência de reações adversas e melhorando a adesão ao tratamento. No presente trabalho, dois métodos bioanalíticos foram desenvolvidos e validados para análise do TPM em amostras de plasma, por eletroforese capilar utilizando os detectores DAD (detecção indireta) e C4D Lite (detecção condutométrica). A análise eletroforética para ambos os métodos foi realizada em um capilar de sílica fundida de 75 ?m de diâmetro interno e 40 cm de comprimento efetivo. O eletrólito de corrida foi composto por trietanolamina (TEA) 15 mmol L-1 para a detecção condutométrica e TEA 15 mmol L-1, ácido 3,5 dinitrobenzóico (DNB) 5 mmol L-1 e brometo de cetiltrimetilamônio (CTAB) 0,4 mmol L-1 para detecção indireta. As injeções foram feitas no modo hidrodinâmico a 0,8 psi por 5 segundos e a tensão aplicada foi de -20 e 20 kV para detecção indireta e condutométrica, respectivamente. Nestas condições, o TPM migrou em 2,5 min. O preparo das amostras foi realizado por extração líquido-líquido (LLE), usando éter metil-terc-butílico (MTBE) como solvente extrator e utilizando 500 e 200 ?L de plasma para o método com detecção indireta e condutométrica, respectivamente. Os métodos foram validados de acordo com os guias oficiais da Agência Nacional de Vigilância Sanitária e European Medicine Agency e apresentaram linearidade no intervalo de concentração plasmática de 1-30 ?g mL-1 e valores de precisão e exatidão abaixo de 15 %, além de serem seletivos. Assim, ambos os métodos foram aplicados com sucesso em amostras de plasma de pacientes em tratamento com o TPM. Embora os detectores universais geralmente apresentem um maior limite de detecção comparado com detectores seletivos, o método com detecção condutométrica desenvolvido neste trabalho apresentou como vantagem maior sensibilidade em relação método indireto (DAD). Além disso, apresentou também vantagens como simplicidade e baixo custo, mostrando viabilidade na aplicação para análises de rotina.
Titre en anglais
Development and validation of method for analysis of topiramate in plasma samples by capillary electrophoresis with UV/vis and C4D detection
Mots-clés en anglais
C4D
Capillary electrophoresis
detector
human plasma
topiramate
UV/vis
Resumé en anglais
Topiramate (TPM) is an antiepileptic drug that has been used as an agent for the control of partial and generalized seizures in both adults and children patients. The determination of the plasmatic concentration of TPM in association with clinical observations is of utmost importance for the individual adjustment of the administered dose to the patient. In the present work, two bioanalytical methods have been developed and validated for TPM analysis in plasma samples by capillary electrophoresis using DAD (indirect detection) and C4D Lite detectors (conductometric detection). The electrophoretic analysis for both methods were performed on a fused silica capillary of 75 ?m of internal diameter and 40 cm effective length. The background electrolyte was composed of triethanolamine (TEA) 15 mmol L-1 for conductometric detection and TEA 15 mmol L-1, 3,5 dinitrobenzoic acid (DNB) 5 mmol L-1 and cetyltrimethylammonium bromide (CTAB) 0.4 mmol -1 for indirect detection. The injections were carried out in hydrodynamic mode, at 0.8 psi for 5 seconds, and the voltage applied were -20 e 20 kV for indirect and conductometric detection, respectively. Under these conditions, TPM migrated in 2.5 min. The sample preparation was carried out by liquid liquid extraction (LLE) using methyl tert-butyl ether (MTBE) as solvent, and using 500 and 200 ?L of plasma samples for direct and conductometric detection, respectively. The methods were validated according to the official guides of the Agência Nacional de Vigilância Sanitária and European Medicine Agency, and showed linearity in plasmatic concentration range of 1-30 mg L-1, precision and accuracy values below 15%, beside selectivity. Both methods have been applied successfully to analysis of plasma samples of patients in treatment with TPM. Although universal detectors have a higher quantification limit compared with selective detectors, the method with conductometric detection developed in this work presented greater sensitivity than indirect method (DAD). Furthermore, it also presented advantages such as simplicity and low cost, showing feasibility in applying to routine analysis.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2016-10-05
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.