Full name
Maraine Catarina Tadini
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
Ribeirão Preto, 2016
Supervisor
Committee
Oliveira, Marcelo Firmino de (President)
Olivi, Paulo
Paixão, Thiago Regis Longo Cesar da
Title in Portuguese
Desenvolvimento de sensores eletroquímicos para a detecção voltamétrica de MDMA em amostras de interesse forense
Keywords in Portuguese
CB[7]
MDMA
Nafion
Química Forense
Sensor
Voltametria
Abstract in Portuguese
A 3,4-metilenodioximetanfetamina (MDMA) é a principal substância psicoativa comercializada ilegalmente em comprimidos de ecstasy. O MDMA é uma droga de ação psicotrópica e uso proscrito, conforme lista F (grupo F2) da ANVISA, pois apresenta propriedades alucinógenas e estimulantes e seu uso/abuso pode gerar uma série de danos à saúde dos usuários. O desenvolvimento de eletrodos quimicamente modificados (EQMs) na eletroanalítica tem por finalidade a obtenção de sistemas de detecção mais sensíveis e seletivos para o analito de interesse. Também, considera-se necessário desenvolver novas técnicas e métodos para a detecção de compostos em amostras de interesse forense, a fim de obter ferramentas para auxiliar os cientistas forenses no combate ao comércio ilícito de substâncias. Conforme problemática exposta, este trabalho teve por finalidade o desenvolvimento de eletrodos quimicamente modificados utilizando como modificadores da superfície eletródica de carbono vítreo o Nafion e Nafion/CB[7], utilizando deposição por drop coating e spin coating para a detecção de MDMA através das técnicas de voltametria cíclica e onda quadrada. Conforme o sistema empregado, os melhores EQMs desenvolvidos foram de Nafion (1,5% v/v) e Nafion (1,5% v/v)/CB[7] (10,0 µg.mL-1). Os EQMs desenvolvidos apresentaram limite de detecção e quantificação na faixa de traços e menores que aqueles reportados em outros trabalhos da literatura. Considerando a aplicação dos EQMs para a detecção de MDMA em amostras de ecstasy, verificaram-se as respostas voltamétricas de outras substâncias: cafeína, metanfetamina, teobromina, lidocaína, cloridrato de procaína, (±)-metanfetamina e cloridrato de cocaína. Nas condições experimentais empregadas, observou-se que as substâncias estudadas não atuam como falsos positivos para o MDMA. Paralelamente, obtiveram-se onze lotes de comprimidos de ecstasy (apreendidos e cedidos pela Polícia técnico-científica de Ribeirão Preto-SP) e realizaram-se análises qualitativas e quantitativas nos mesmos, utilizando técnicas colorimétricas (Marquis, Ácido sulfúrico, Simon e Simon com acetona) e cromatográficas (CG-EM E CLAE-EM). Considerando o melhor EQM desenvolvido, quantificaram-se 11 lotes de ecstasy pela técnica voltamétrica e cromatográfica, dentre os lotes estudados, dois não continham MDMA, um apresentou uma mistura de MDMA e cafeína e os demais continham MDMA. A concentração de MDMA presente nos lotes variou de 0 até 61 % em massa. A detecção de MDMA em ecstasy pelo método voltamétrico desenvolvido se mostrou viável e sensível para o analito de interesse.
Title in English
Development of electrochemical sensors for voltammetric detection of MDMA in samples of forensic interest
Keywords in English
CB[7]
Forensic chemistry
MDMA
Nafion
Sensor
voltammetry
Abstract in English
The 3,4-methylenedioxymethamphetamine (MDMA) is the main psychoactive component of ecstasy tablets, that have an illicit trade. MDMA has been an illicit psychotropic drug, and it has a prohibited use (group F2, in ANVISAs F list), because of its hallucinogenic and stimulating effects, and the use/abuse can poses serials health risks. The development of chemically modified electrodes (CME) in electroanalytical methods aims to get more sensitive and selective systems to detect the analytes. In this context, it is necessary to develop new techniques and methodologies to the detection of illicit samples; it provides more tools to help the forensic scientists to combat the illicit drug trade. So, this work focused in the development of chemically modified electrodes (CMEs) with modifications on the glassy carbon surface by drop coating and spin coating using Nafion and Nafion/CB[7] solutions. The CMEs were tested using cyclic, and square wave voltammetry to detect MDMA. Considering the employed system, the best CMEs were made by Nafion (1.5% v/v), and Nafion (1.5% v/v)/CB[7] (10.0 µg.mL-1) thin films. It was possible to observe better sensitivities for these sensors, in comparison to other MDMA studies reported in the literature. The specificity of the proposed sensors was checked in relation to other drugs: caffeine, methamphetamine, theobromine, lidocaine, procaine hydrochloride, and cocaine hydrochloride. These drugs do not interfere in this voltammetric method. Additionally, we studied eleven lots of ecstasy samples, allowed by the Scientific Police - Ribeirão Preto-SP, and we provide qualitative and quantitative studies using colorimetric techniques (Marquis, Sulfuric acid, Simon, and Simon with acetone), and chromatografic techniques (GC-MS and HPLC-MS). The MDMA quantification in real samples was obtained by high performance liquid chromatography with a mass spectrophotometer, and we compared with the voltammetric technique, using the developed CME. Between the analyzed lots, two of them didnt present in their composition, one lot had a mix of caffeine and MDMA, and another presented MDMA. The MDMAs concentration in lots had a large variation, with 0 to 61 % w/w. The MDMAs voltammetric detection in ecstasy lots was viable. And, it is also possible to apply this methodology to analyze MDMA traces.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2016-12-12
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.