Master's Dissertation
DOI
https://doi.org/10.11606/D.59.2015.tde-26022015-104619
Document
Author
Full name
Alessandra Maia Marques Martinez Perez
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
Ribeirão Preto, 2015
Supervisor
Committee
Poletti, Martin Eduardo (President)
Medeiros, Regina Bitelli
Tomal, Alessandra
Title in Portuguese
Estudo experimental da otimização em sistemas de mamografia digital CR e DR
Keywords in Portuguese
Figura de mérito
Mamografia digital
Otimização técnica
Qualidade de imagem
Redução de dose
Abstract in Portuguese
A recente inserção e forte avanço da mamografia digital no Brasil como ferramenta de rastreamento do câncer mamário e as evidências de outras condições de otimização, quando comparadas à mamografia convencional (tela filme), requerem que novos parâmetros de qualidade sejam incluídos e estudados, bem como que as condições de otimização sejam revistas. O objetivo deste trabalho foi determinar a técnica radiográfica otimizada para dois sistemas de detecção (CR e DR) em uso em três unidades de mamografia: Mammomat 3000 Nova (Siemens), Senographe DMR (GE) e Senographe 2000D (GE). A otimização foi conduzida para uma variedade de combinações de fatores técnicos e configurações de simuladores de mama, tais como valores de kilovoltagem (26 a 32 kV), combinações anodo/filtro (Mo/Mo, Mo/Rh e Rh/Rh), material simulador de mama de várias espessuras (2 a 8 cm) e lesões simuladas como massas e calcificações, usando uma figura de mérito (FOM) como parâmetro. Verificou-se que o uso da combinação anodo/filtro que gera os espectros mais energéticos em cada equipamento proporcionou os maiores valores de FOM para todas as espessuras de simulador de mama e voltagens, devido a redução da dose. As combinações anodo/filtro que deram esses resultados foram Mo/Rh para o equipamento da marca Siemens e Rh/Rh para ambos os equipamentos da marca GE, correspondentes aos espectros mais energéticos de cada unidade. Foi observada ainda uma tendência de aumento do kV que maximiza FOM com o aumento da espessura.
Title in English
Experimental study of optimization in CR and DR digital mammography systems
Keywords in English
Digital mammography
Dose reduction
Figure of merit
Image quality
Technique optimization
Abstract in English
The recent introduction and intense advance of digital mammography in Brazil as a tool in breast cancer screening and the evidences of new optimization conditions when compared to conventional mammography (screen-film) require adding and studying novel quality parameters, as well as revisiting optimization conditions. The objective of this work was to determine optimized radiographic technique for two detection systems (CR and DR) in use in three mammography units: Mammomat 3000 Nova (Siemens), Senographe DMR (GE) and Senographe 2000D (GE). Optimization was conducted for various combinations of technique factors and breast phantom configurations, such as kilovoltage settings (26 to 32 kV), target/filter combinations (Mo/Mo, Mo/Rh and Rh/Rh), breast equivalent material in various thicknesses (2 to 8 cm) and simulated mass and calcification lesions, using a figure of merit (FOM) as a parameter. When using anode/filter combination which generates higher energy spectra in each equipment, it was verified that higher FOM values were achieved for all voltages and phantom thicknesses, due to dose reduction. Anode/filter combinations which led to those results were Mo/Rh for Siemens equipment and Rh/Rh for both GE equipments, corresponding to the higher energy spectra in each unity. It was also observed an increasing tendency of kV which maximizes FOM with the increase of thickness.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2015-05-08
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.