• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Disertación de Maestría
DOI
https://doi.org/10.11606/D.59.2014.tde-24072014-150954
Documento
Autor
Nombre completo
Antonio Carlos Nunes Bertolo
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
Ribeirão Preto, 2014
Director
Tribunal
Moraes, Eder Rezende (Presidente)
Costa, Alessandro Martins da
Silva, Ana Maria Marques da
Título en portugués
Avaliação do desempenho de um sistema de tomografia PET com geometria elipsoidal
Palabras clave en portugués
NECR e Resolução Espacial.
Sensibilidade do PET
sistema PET elíptico
Tomografia por Emissão de Pósitrons
Resumen en portugués
Dentre as técnicas diagnósticas por imagem, em Medicina Nuclear, têm-se destacado a Tomografia por Emissão de Pósitrons (PET). O PET fornece imagens funcionais da região ou órgão de interesse, possibilitanto o diagnóstico de várias doenças e, também, um mapeamento da evolução ou regressão das mesmas. Os atuais sistemas PET apresentam blocos de cristais arranjados em geometria circular, mas a busca por novos arranjos, em geometrias diferenciadas, é pertinente, pois pode possibilitar um melhor desempenho destes tomógrafos. O objetivo deste estudo é avaliar o desempenho de um sistema PET constituído por blocos de cristais em arranjo elipsoidal. Para tal estudo, fez-se uso do GATE, ambiente de simulação para tomografia por emissão que apresentou resultados satisfatórios comparados à aparelhos utilizados na rotina clínica. Desta forma, elaborou-se dois sistemas PET, o primeiro com 46 blocos de cristais arranjados em geometria circular, no qual, os detectores opostos foram separados a uma distância de 816,4 mm na direção transaxial. No segundo caso, utilizou-se 36 blocos de cristais, arranjados em geometria elíptica, onde os detectores foram separados a uma distância de 500 mm na direção vertical e 816,4 mm na direção horizontal. Uma vez realizadas as simulações, fez-se os testes de validação para Sensibilidade, Fração de Espalhamento, NEC (Noise Equivalente Count Rate) e Resolução Espacial, para então, avaliar e comparar o desempenho do sistema PET para ambas as geometrias propostas. Estes testes foram feitos de acordo com as especificações da norma NEMA NU 2-2007. Para o teste de sensibilidade, com o phantom localizado no centro do FOV transaxial do tomógrafo, obteve-se um ganho no sistema PET, com geometria elíptica, de 28,7% em relação ao sistema convencional, de geometria circular. Para o phantom deslocado à 10 cm do centro do FOV transaxial, a sensibilidade do sistema PET elíptico foi 26,2% superior em relação ao circular. O tempo de aquisição para cada simulação neste teste foi de 400 s. Para os testes de Fração de Espalhamento e NEC foram realizadas 33 simulações, para cada geometria, variando o tempo de aquisição e a atividade. A fração de espalhamento na geometria elíptica foi de 35,5% e na circular 34,6%. Na geometria circular, a curva NEC atinge o pico máximo em 259,3 kcps a uma concentração de atividade de 34,1 kBq ml-1 , já para geometria elíptica, o pico máximo é atingido em 239,1 kcps a uma concentração de atividade de 24,8 kBq ml-1 . As medidas foram adquiridas em três posições transaxiais (x,y) no centro do FOV e, posteriormente, deslocadas em ¼ do centro do FOV, totalizando 6 medidas para cada geometria. Para cada uma destas 6 medidas, encontrou-se a FWHM (largura a meia altura) e a FWTM (largura a 10% do máximo) nos 3 eixos de coordenadas (x,y,z), assim, para cada geometria proposta obteve-se 18 valores de FWHM e 18 valores de FWTM. A geometria elíptica apresentou melhoras na resolução espacial em FWHM em 4 dos 18 valores de resolução. Já para FWTM, a geometria elíptica apresentou melhora em 8 dos 18 valores de resolução. Em relação a FWHM, onde houve melhora na resolução espacial, teve-se um ganho médio de 18,7%. Já as perdas, em resolução espacial foi em média 43,26%. Para FWTM, a melhora em resolução espacial foi em média 12,09%, mas a perda de resolução, neste caso, foi em média 45,59%. Pode-se concluir que a geometria proposta apresenta algumas vantagens em relação a geometria convencional. Na geometria elíptica há uma redução de 10 blocos de cristais em relação a geometria circular, o que possivelmente gera uma redução no custo de fabricação do tomógrafo. Em relação a sensibilidade, há uma melhora significativa para nova geometria, bem como uma redução da concentração da atividade na qual o tomógrafo alcança seu melhor desempenho, embora ocorra uma ligeira diminuição na taxa de contagens no sistema elíptico. Essa redução na concentração de atividade pode ocasionar uma redução no tempo de exame e utilização de menos material radioativo, reduzindo o risco ao paciente e o custo do exame, bem como possibilitando a realização de um maior número de exames num mesmo intervalo de tempo. A fração de espalhamento do sistema elíptico é ligeiramente maior em relação ao sistema circular. A principal perda do sistema PET elíptico está na resolução espacial, onde não houve melhoras para maior parte das medidas realizadas, necessitando de alguns ajustes que pode se dar, entre outras coisas, no arranjo dos blocos de cristais. Acredita-se que ajustes na angulação dos detectores e na formação das coincidências seja possível melhorias nas resoluções, e NEC.
Título en inglés
Performance evaluation of a system for PET tomography ellipsoidal geometry.
Palabras clave en inglés
Eliptical PET system
NECR and Spatial Resolution.
Positron Emission Tomography
Sensitivity of PET
Resumen en inglés
Among the diagnostic imaging techniques, in Nuclear Medicine, have been prominent in Positron Emission Tomography (PET). PET provides functional images of the region or organ of interest, allowing diagnostic of various diseases and also a mapping of the development or regression of the same. Current PET systems feature blocks arranged in circular geometry crystals, but the search for new arrangements, in different geometries, is relevant because it can provide a better performance of these scanners. The objective of this study is to evaluate the performance of a PET system consisting of blocks of crystals ellipsoidal arrangement. For this study, we used the GATE simulation environment for emission tomography that presented satisfactory results compared to the devices used in the clinical routine. This way we prepared two PET systems, the first one containing 46 blocks of crystals arranged in a circular geometry, where the opposite detectors were separated by a distance of 816.4 mm in the transaxial direction. In the second case, we used 36 blocks of crystals arranged in an elliptical geometry, where the detectors are separated a distance of 500 mm in the vertical direction and 816.4 mm in the horizontal direction. After the simulations were performed, the validation tests for sensitivity, scatter fraction, NEC (Noise Equivalent Count Rate) and Spatial Resolution were made, to finally evaluate and compare the performance of the PET system for both proposed geometries. The tests were made accordingly to specified NEMA NU 2-2007 standard. For the sensitivity test, with the phantom located in the center of FOV transaxial of the tomography, we obtained a gain in PET system with elliptical geometry of 28.7% compared to the conventional system, the circular geometry. For the phantom shifted to 10 cm from the center of the transaxial FOV, the sensitivity of PET elliptical system was 26.2% higher than in the circular. The acquisition time for each simulation in this test was 400 s. For Scatter Fraction test and NEC, were made 33 simulations for each geometry, varying the time of acquisition and activity. The scatter fraction in the elliptical geometry was 35,5% and 34,6% in the circular geometry. In the circular geometry, NEC curve reaches its peak in 259.3 kcps at an activity concentration of 34.1 kBq ml-1, for the elliptical geometry, the maximum peak is reached at 239.1 kcps at an activity concentration of 24.8 kBq ml-1. For the spatial resolution test we used a point source containing 18F with 4 MBq activity and acquisition time of 200 s. The measurements were acquired in three transaxials positions (x,y) in the center of FOV and, shifted in ¼ of the center of the FOV, summarizing 6 measurements for each geometry. For each one of these 6 measurements, we found the FWHM (Full width at half-maximum amplitude) and the FWTM (Full width at tenth-maximum amplitude) in the three axis (x,y,z), in this way, for each proposed geometry, we obtained 18 values of FWHM and 18 values of FWTM. The elliptical geometry showed improvement in the spatial resolution in FWHM in 4 of the 18 resolution values. For the FWTM, the elliptical geometry showed improvement in 8 of 18 resolution values. Regarding the FWHM, where there was an improvement in spatial resolution, we obtained the average gain of 18,7%. For the losses, in spatial resolution was an average of 43,26%. For the FWTM, the improvement in spatial resolution was an average percentage of 12,09%, while the loss of resolution had the average percentage of 45,59%. It can be concluded that the proposed geometry showed some advantages regarding the traditional geometry. In the elliptical geometry there is one reduction of 10 blocks of crystals in relation to the circular geometry, which possibly can cut fabrication costs of the tomography. Regarding to sensibility, there is a significant improvement for the new geometry, as well as one reduction of the activity concentration in which the tomography reaches the best performance, although the occurrence of a slight fall in the counts in the elliptical system. This reduction in the activity concentration can help to reduce the test time and the use less radioactive material, reducing the risk for the patients and the test costs, as well as improving the number of tests in the same period of time. The scattering fraction in the elliptical system is slight higher to the circular system. The major loss in the elliptical PET system is the spatial resolution, where there was not improvement for the great fraction of the obtained measurements, requiring some modifications that can be made, in special, in the crystal blocks arrangement. It is believed that some modifications to the angle of the detectors and in the formation of the coincidences can improve the resolutions and NEC.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2014-09-19
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.