Doctoral Thesis
DOI
https://doi.org/10.11606/T.59.2014.tde-23052014-104532
Document
Author
Full name
Francisco Glaildo Almeida Sampaio
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
Ribeirão Preto, 2014
Supervisor
Committee
Almeida, Adelaide de (President)
Appoloni, Carlos Roberto
Caldas, Linda Viola Ehlin
Caliri, Antonio
Goes, Evamberto Garcia de
Title in Portuguese
Avaliação de gap em feixes de fótons e elétrons na radioterapia utilizando o dosímetro Fricke Xilenol Gel e sistema de leitura CCD
Keywords in Portuguese
Campos Adjacentes
Fricke Xilenol Gel
Gap
Método Monte Carlo.
Abstract in Portuguese
Em radioterapia, dependendo do tamanho do volume alvo a ser irradiado, pode ser necessário mais de um feixe de radiação (fótons ou elétrons). Isto faz com que sobreposições dos feixes, tanto na superfície (campos de radiação) quanto em profundidade (composição das penumbras) resultem em sobredosagens que podem interferir nos tratamentos e restabelecimento dos pacientes. Para se evitar a sobredosagem foram criados os gaps (espaçamento ou distanciamento entre os campos dos feixes, na superfície de entrada dos mesmos) com a finalidade de diminuir a sobredosagem e homogeneizar a dose absorvida dentro de valores aceitáveis, que não devem exceder 20% acima do valor esperado. Para feixes de elétrons não há a prática do uso de gap, como para feixes de fótons, e não há equações que possam prever a sobredosagem na adjacência desses feixes, assim como não há um estudo mais aprofundado das técnicas possíveis de serem utilizadas para se evitar a sub e/ou sobredosagem desses feixes na superfície e em profundidade. Com isto, este trabalho foi estruturado visando à obtenção da homogeneidade das doses absorvidas na região de adjacência dos feixes de elétrons, partindo do conhecimento da aplicação de gaps para feixes de fótons. Para tal, foi desenvolvido um protótipo de leitura dos dosímetros FXG e filme radiocrômico, baseado no sensor Charge-Coupled Device (CCD). Esse protótipo permitiu fazer as medidas de diversos feixes de fótons e elétrons, variando suas energias, geometrias e superfícies de entrada (plana e/ou curvas), cujas respostas puderam ser comparadas com as obtidas por simulação por Método Monte Carlo (MMC), resguardando as mesmas condições de feixes e arranjos experimentais. Dos resultados obtidos, pode-se comprovar: a eficácia do protótipo desenvolvido, o desenvolvimento de uma equação para prever a sobredosagem advinda da adjacência de feixes de elétrons a partir de dados experimentais, a reversão do uso da equação de gap específica para feixes de fótons para feixes de elétrons em superfícies planas, a equivalência dos resultados da simulação por MMC com os experimentais de gap realizados, além da adequabilidade dos filtros compensadores (em cunha) desenvolvidos para superfícies de entrada curvas. Todos esses resultados indicam a possibilidade da utilização do gap para feixes adjacentes de elétrons em superfícies planas e filtros em cunha em superfícies curvas, passíveis de serem utilizados na prática radioterápica.
Title in English
Evaluation of gap for photon and electron beams in radiotherapy through the use of Fricke Xylenol Gel dosimeter and CCD reading system
Keywords in English
Adjacent Fields
Fricke Xylenol Gel
Gap
Monte Carlo Method.
Abstract in English
Radiotherapy of extensive volumes normally is made using more than one radiation beam (photons or electrons). These beams once are divergent, in the intersection of their penumbra the isodose composition value is increased increasing the absorbed dose value in the treatment region. To overcome this effect is used the \textit (a shift between the beam fields at the entrance surface) whose scope is to homogenize the depth absorbed dose and do not let it to exceed 20% of its expected value. Only for photons exists the practice of gap (with equations predicting gap for plane surfaces) and compensator filters (for curve surfaces). Once these ideas are non existent for electron adjacent beams and once the use of this type of radiation has increased along the years for patient irradiation, most of the times replacing the photon beams, the curiosity to obtain methods to minimize the overdose due to the electron adjacent beams was triggered and turn it the goal of this research. For this work two dosimeters were used (Fricke Xylenol Gel, FXG and Radiocromic film) and their measurements were made through a developed prototype with a Charge-Coupled Device, CCD, whose results were compared also with Monte Carlo simulations considering the same experimental conditions (irradiation geometry, beam energy, entrance surface geometry, plane and/or curve, and setup). From the results obtained one could notice that: the CCD prototype resulted in an adequate device to be used with the selected dosimeters, the gap equation for photons could be used successfully also for adjacent electron beams on plane surfaces, compensator filters (acrylic wedge filters) are adequate for curve surfaces and the Monte Carlo Simulation results shown equivalence with the experimental ones. All these results encourage the gap for electron adjacent beams on plane surfaces, as well the use of wedge filters for curve surfaces in radiotherapy practice.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2014-09-16
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.