Master's Dissertation
DOI
https://doi.org/10.11606/D.59.2009.tde-19042010-175558
Document
Author
Full name
Mairon Marques dos Santos
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
Ribeirão Preto, 2009
Supervisor
Committee
Nicolucci, Patrícia (President)
Calvo, Wilson Aparecido Parejo
Siqueira, Paulo de Tarso Dalledone
Title in Portuguese
Estudo de uma câmara de ionização tipo poço através de simulação Monte Carlo
Keywords in Portuguese
câmara poço
medicina nuclear
PENELOPE
simulação Monte Carlo
Abstract in Portuguese
O uso de simulação Monte Carlo do transporte de radiação na matéria tem sido cada vez mais empregado nas áreas de física radiológica e dosimétrica. Em Medicina Nuclear é possível usar diversos códigos de simulação como ferramenta para estudar diferentes características de resposta de calibradores de dose. O código de simulação Monte Carlo PENELOPE (Penetration and ENErgy LOss of Positron and Electrons) possui um algoritmo misto do transporte de radiação, condensando eventos de interação, conforme os parâmetros de entrada estabelecidos. Neste trabalho, o código de simulação PENELOPE foi usado para estudar a resposta de uma câmara de ionização em função de diversos parâmetros que influenciam suas características de resposta. Neste trabalho os parâmetros de resposta de uma câmara de ionização tipo poço foram estudados através de simulação Monte Carlo, comparando-se os resultados obtidos com dados experimentais. A eficiência da câmara foi testada através de simulação e mostrou-se condizente com os valores previstos através de cálculos. Com relação à atividade, a resposta se mostrou linear para todos os nuclídeos estudados, sendo possível se obter a sensibilidade relativa da câmara através de simulação e de medidas experimentais. A resposta da câmara em função a energia, obtida através de simulação, também representou bem os valores experimentais, sendo possível extendê-los para energias mais altas e mais baixas que os valores das medidas experimentais. Estudos do volume de radiofármaco e da posição da fonte no poço da câmara obtidos através de simulação apresentaram comportamento esperado de acordo com a literatura. O código PENELOPE foi validado para o estudo desta câmara de ionização, permitindo que parâmetros geométricos e de materiais sejam estudados sem os custos e as dificudades dos arranjos experimentais.
Title in English
Study of a well-type ionization chamber by Monte Carlo simulation
Keywords in English
Monte Carlo simulation
nuclear medicine
PENELOPE
well-type ionization chamber
Abstract in English
The use of Monte Carlo simulation to the radiation transport in matter has been widly applied in the radiological and dosimetric areas. In Nuclear Medicine it is possible to use a variety of simulation codes as tools to study different response characteristics of dose calibrators used to measure radionuclides activities. The PENELOPE (Penetration and ENErgy LOss of Positron and Electrons) Monte Carlo simulation code has a mixed algorithm for the transport of radiation, which condenses the interaction events according to the input parameters. In this work, the PENELOPE code of simulation was used to study the response of an ionization chamber as function of parameters influencing its response. The chamber efficiency was tested by simulation and it showed a good agreement with calculated results. To the activity, its response showed a linear behavior for all studied nuclides, allowing one to obtain its sensitivity by simulation and measurements. The response of the chamber as a function of the energy obtained by simulation also showed a good agreement with the measurements, allowing one to extrapolate it to energies below and above the measured ones. The analysis with the volume of radiopharmaceuticals and position of the sourse in the chamber well obtained by simulation showed the expected behavior compared to the ones in literature. PENELOPE was validated to study this ionization chamber, so allow one to perform geometric and material parameters studies without experimental costs and difficulties.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2010-08-23
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.