Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.59.2018.tde-18012018-152142
Document
Auteur
Nom complet
Daiane Miron de Souza
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
Ribeirão Preto, 2017
Directeur
Jury
Costa, Alessandro Martins da (Président)
Neves, Lucio Pereira
Neves, Ubiraci Pereira da Costa
Potiens, Maria da Penha Albuquerque
Titre en portugais
Reconstrução espectral de feixes de raios X diagnósticos
Mots-clés en portugais
Atenuação
Espectrometria
Problema Inverso
Radiodiagnóstico
Raios X
Resumé en portugais
A caracterização completa de um feixe de raios X diagnósticos baseia-se na medição de seu espectro de fluência. O espectro de fluência pode ser medido diretamente utilizando métodos espectroscópicos, no entanto, exige equipamento especializado e não é facilmente realizável em ambiente clínico. Neste trabalho foi implementada uma metodologia que permite obter o espectro de raios X de forma indireta. Esta metodologia baseia-se na aplicação de um modelo matemático que utiliza a transformada inversa de Laplace da curva de atenuação do feixe, gerando dados sobre a distribuição espectral deste. As curvas de atenuação foram medidas por meio de uma câmara de ionização e filtros de alumínio de alta pureza. Os espectros reconstruídos foram validados por meio da comparação da fluência em energia calculada a partir destes e a fluência em energia calculada a partir das curvas de atenuação obtidas experimentalmente. Como aplicação também foram calculados alguns dados característicos do feixe, como primeira e segunda camada semirredutora, energia média, 10º percentil e fator de kerma. Os resultados da fluência em energia calculada a partir dos espectros reconstruídos e da fluência em energia calculada a partir das curvas de atenuação obtidas experimentalmente apresentaram boa concordância, validando os espectros reconstruídos e mostrando o valor da análise da curva de atenuação em comparação com métodos espectroscópicos uma vez que os dados de atenuação podem ser obtidos com comparativa facilidade. Os dados característicos do feixe calculados a partir dos espectros obtidos neste trabalho apresentaram resultados satisfatórios, como era de se esperar, uma vez que a integração é um processo de regularização para a distribuição espectral
Titre en anglais
Spectral Reconstruction of Diagnostic X-Ray Beams
Mots-clés en anglais
Attenuation
Inverse Problem
Radiodiagnosis
Spectrometry
X-ray
Resumé en anglais
A complete characterization of a diagnostic X-ray beam is based on the measurement of its fluency spectrum. The fluency spectrum can be measured directly using spectroscopic methods, however, require specialized equipment and is not easily performed in a clinical environment. In this work a methodology was implemented that allows to obtain the X-ray spectrum in an indirect way. This methodology is based on the application of a mathematical model that uses an inverse Laplace transform of the beam attenuation curve, generating data on a spectral distribution of this beam. The attenuation curves are measured by means of an ionization chamber and high purity aluminum filters. The reconstructed spectra were validated by means of the comparison of energy fluency calculated from them and an energy fluence calculated from the experimentally obtained attenuation curves. As an application, some characteristic beam data were also calculated, such as first and second half-value layers, mean energy, 10th percentile and kerma factor. The energy fluency calculated from the reconstructed spectra and the energy fluence calculated from the experimentally obtained attenuation curves showed good agreement, validating the reconstructed spectra and showing the value of the attenuation curve analysis in comparison with one spectroscopic methods attenuation data can be obtained with comparative ease. The characteristic beam data calculated from the spectra obtained in this work presented satisfactory results, as expected, since it is a regularization process for a spectral distribution
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
DissertCorrigida.pdf (1.16 Mbytes)
Date de Publication
2018-04-23
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.