Doctoral Thesis
DOI
https://doi.org/10.11606/T.59.2010.tde-17082010-160317
Document
Author
Full name
Jose Luis Javier Vega Ramirez
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
Ribeirão Preto, 2010
Supervisor
Committee
Baffa Filho, Oswaldo (President)
Ghilardi Netto, Thomaz
Nicolucci, Patrícia
Oliveira, Harley Francisco de
Pina, Diana Rodrigues de
Title in Portuguese
Estudos dosimétricos em interfaces teciduais em radioterapia utilizando dosimetria por ressonância paramagnética eletrônica EPR
Keywords in Portuguese
alanina
EPR
minidosímetros
objeto simulador heterogênea.
PDP
PENELOPE
Abstract in Portuguese
A acurácia e precisão exigida nas novas técnicas radioterápicas é bastante elevada, como no caso da Radioterapia com Intensidade Modulada (IMRT). Desta forma, o tamanho dos dosímetros a serem utilizados nas medidas experimentais é um fator crítico. Além disso, a ausência de equilíbrio eletrônico lateral e a existência de intensos gradientes de dose no campo apresentam um desafio na determinação da dose administrada no tumor. Alguns dos dosímetro utilizados atualmente não apresentam resolução espacial ou sensibilidade condizentes com os requisitos das dosimetrias clínicas nessas técnicas radioterápicas. A dosimetria por Ressonância Paramagnética Eletrônica (EPR), utilizando o aminoácido L-alanina como dosímetro, é um método amplamente aceito de dosimetria para medições de altas doses de radiação. O objetivo desta tese foi aprofundar o desenvolvimento da dosimetria por EPR, como um método competitivo para aplicações na radioterapia, que preencha os requisitos de precisão e acurácia e comparar medidas realizadas com os minidosímetros de L-alanina/ EPR com outras técnicas e algoritmos estudados. A nova metodologia dos minidosímetros com 95% L-alanina e 5% PoliVinil Álcool (PVA) de 1 mm de diâmetro e 3 mm de altura com massa 3,5 mg a 4,0 mg., foi utilizadas para medições de porcentagem de dose profunda (PDP) e perfis de campo de irradiação em objetos simuladores físicos (objetos simuladores não homogêneos) irradiados com a técnica da radioterapia. Os objetos simuladores não homogêneos foram construídos com materiais equivalentes que simulam o tecido mole, pulmão, osso e outras estruturas do corpo humano como o titânio, que é utilizado em próteses. Os objetos simuladores foram irradiados com fótons de energia máxima de 6 MeV do acelerador linear Siemens Primus com uma dose de 20 Gy aproximadamente. Os resultados obtidos experimentalmente com os minidosímetros de L-alanina foram comparados com outros procedimentos e técnicas já validadas como, por exemplo, o filme, o sistema de planejamento do tratamento (SPT) e a simulação Monte Carlo com o código PENELOPE-2008, que possibilitou a comparação de dados experimentais com dados obtidos pelo sistema de planejamento, e análise do comportamento das doses na interface de diversos materiais equivalentes a tecidos. As curvas de porcentagem de dose profunda (PDP) simuladas com o PENELOPE foram validadas pelas PDPs do setor de Radioterapia do HCFMRP-USP e utilizadas como padrão dosimétrico para nosso estudo. A menor diferença percentual entre a dose absorvida no caso simulado comparado ao experimental foi de 0,12%, e a maior de 7,2%. O material de maior absorção de dose foi o Titânio com um fator de atenuação de 40,43%. Os resultados obtidos neste trabalho demonstram que o sistema dosimétrico utilizando EPR com minidosímetros de 95% L-alanina e 5% PVA apresentam grande potencial na determinação de dose nas aplicações das novas técnicas de Radioterapia de pequenos campos.
Title in English
Studies on Tissue Interfaces Dosimetry in Radiotherapy Using Electron Paramagnetic Resonance Dosimetry by EPR
Keywords in English
alanine
EPR
inhomogeneous phantom and PENELOPE
minidosimeters
PDD
Abstract in English
The accuracy and precision required in new radiation techniques is quite high, as is the case of Intensity Modulated Radiotherapy (IMRT). Thus, the size of the dosimeters to be used in experimental measurements is critical. Moreover, the absence of lateral electronic equilibrium and the existence of intense dose gradients in the field present a challenge in determining the dose in the tumor. Some of the dosimeter used nowadays do not have spatial resolution or sensitivity consistent with the requirements of clinical dosimetry in these radiation techniques. The Electron Paramagnetic Resonance (EPR) dosimetry using the amino acid L-alanine as a dosimeter is a widely accepted method of dosimetric measurements of high-dose radiation. The goal of this thesis was to deepen the development of EPR dosimetry as a competitive method for applications in radiotherapy, fulfilling the requirements of precision and accuracy and compare measurements performed with L-alanine / EPR minidosimeters with other techniques and algorithms investigated. The new methodology of minidosimeters with 95% L-alanine and 5% polyvinyl alcohol (PVA), of 1 mm in diameter and 3 mm in height and with a mass 3.5 mg to 4.0 mg, was used for measurements of percentage depth dose (PDD) and field profiles in phantoms (inhomogeneous phantoms) irradiated with radiotherapy technique. The inhomogeneous phantoms were constructed with materials that simulate the soft tissue, lung, bone and other structures of the human body such as titanium, which is used in prostheses. The phantoms were irradiated with photons of 6 MeV maximum energy from a Siemens Primus linear accelerator with a dose of approximately 20 Gy. The results obtained experimentally with the L-alanine minidosimeters were compared with other procedures and techniques already validated such as, for example, film, treatment planning system (TPS) and Monte Carlo simulation with the code PENELOPE-2008, which allowed data comparison with the treatment planning system and an analysis of doses in the interface of different materials equivalent to the tissues. The curves of percentage depth dose (PDP) simulated with PENELOPE were validated with data from the Sector of Radiotherapy of HCFMRP-USP and were used as a dosimetric standard in this study. The lowest difference between simulated and experimental doses was 0.12% and the higher was 7.2%. The material with larger absorption was the Titanium with an attenuation factor of 40.43%. The present results demonstrate that the dosimetric system using EPR minidosimeters with 95% L-alanine and 5% PVA has a great potential in determining the dose from small fields new techniques in radiotherapy.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Ramirez_tese2010.pdf (6.14 Mbytes)
Publishing Date
2010-11-09
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.