Disertación de Maestría
DOI
https://doi.org/10.11606/D.59.2010.tde-17042010-115247
Documento
Autor
Nombre completo
Raul Ernesto Camargo Mendoza
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
Ribeirão Preto, 2010
Director
Tribunal
Poletti, Martin Eduardo (Presidente)
Costa, Alessandro Martins da
Furquim, Tânia Aparecida Correia
Título en portugués
Determinação de grandezas dosimétricas de interesse em mamografia usando detectores termoluminescentes
Palabras clave en portugués
Dose de entrada na pele
Dose em profundidade
Fator de retroespalhamento
Resumen en portugués
Os órgãos de saúde internacionais e nacionais, como o Ministério da Saúde na portaria 453/98 da Vigilância Sanitária, exigem que a Dose de Entrada na Pele seja avaliada para cada equipamento mamográfico através da leitura de um sistema câmara de ionização-eletrómetro corrigida pelo fator de retroespalhamento. Ao não existir menção explícita na portaría de valores utilizáveis para o fator de retroespalhamento, este trabalho visa à determinação experimental do fator de retroespalhamento, através da utilização dos dosímetros termoluminescentes TLD-100. No estudo são verificadas as dependências geométricas e espectrais do fator de retroespalhamento, assim como do valor da Dose de Entrada na Pele, e da Dose em Profundidade, correspondentes com as técnicas radiográficas empregadas nos exames mamográficos convencionais de rotina. Foram avaliados feixes na faixa de 0,35 mmAl até 0,43 mmAl, tensões do tubo de 25kV, 28kV, 30kV, e 32kV, assim como os três tamanhos de campo disponíveis no Mamógrafo Senographe DMR utilizado, e distancias focofilme iguais a 56cm, 61cm e 66cm. Os resultados obtidos foram comparados com publicações existentes, as quais apresentam resultados obtidos através de Simulação Monte Carlo, câmaras de ionização, e dosímetros TLD-100. Os resultados obtidos neste trabalho permitem estabelecer e discutir as dependências das grandezas dosimétricas estudadas com a Camada Semi-Redutora, tensão do tubo, combinação ânodo-filtro, tamanho de campo, distância foco-filme e espessura da mama.
Título en inglés
Determination of dosimetric quantities of interest in Mammography using thermoluminescent detectors.
Palabras clave en inglés
backscatter factor
entrance-skin dose
in-depth dose
TLD dosimeter
Resumen en inglés
National and international health organizations such as the Brazilian Ministry of Health, through its Secretary of Health Surveillance establishes in the publication Nº 453/98 that in all mammographic equipments must be evaluated the entrance-skin dose through the readings of an ionization chamber-electrometer system corrected by the backscatter factor, among others factors. Nevertheless, there is no explicit mention for useful values of backscatter factor in this document; the main aim of this work is the experimental determination of backscatter factor through the use of TLD-100 dosimeters. In this study, the geometric and spectral dependencies of the backscatter factor, entrance-skin dose and the in-depth dose were evaluated, corresponding to the most radiographic techniques employed in conventional mammographic procedures, i.e., beam qualities in the range of 0.35 mmAl to 0.43 mmAl, tube voltages from 25kV to 32kV, focus-film distances from 56cm to 66cm, and three field sizes were evaluated. Our results were compared with those previously published obtained through Monte Carlo simulation, ionization chambers and TLD dosimeters. The results obtained in this work allow studying the dependency of the mentioned dosimetric quantities with the half-value layer, tube voltage, anode-filter combination, field size, focusfilm distance and breasting thickness of the breast.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
a.pdf (3.76 Mbytes)
Fecha de Publicación
2010-04-19
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.