Título en portugués
Otimização de imagens e proteção radiológica em fluoroscopia
Palabras clave en portugués
fluoroscopia
intervencionista
Otimização
otimização.
Resumen en portugués
Este trabalho teve como objetivo avaliar a proteção radiológica em procedimentos fluoroscópicos intervencionistas através de testes de controle de qualidade e levantamento radiométrico, bem como a implementação de um objeto simulador para avaliação da qualidade da imagem e sua otimização, com conseqüente redução da dose no paciente. O objeto simulador, desenvolvido para avaliação de equipamentos de radiologia convencional, mostrou-se eficaz na avaliação da qualidade da imagem em fluoroscopia, bem como na determinação dos níveis de taxa de kerma no ar na entrada da pele de um paciente adulto normal. A avaliação dos parâmetros técnicos dos diversos modos de exposição de um equipamento fluoroscópico digital permitiu estabelecer os indicadores de dose para otimização das imagens em procedimentos intervencionistas. Esta otimização possibilita reduzir a dose no paciente mantendo a qualidade da imagem e, conseqüentemente, sem perda de informação diagnóstica.
Título en inglés
Optimizing Images and Radiological Protection in Fluoroscopy
Palabras clave en inglés
Fluoroscopy
Optimization and intervencionist.
Resumen en inglés
This work had as main objective to evaluate the radiological protection in the interventional procedures radiology through quality control and evaluations radiometric, as well as the implementation of a phantom for evaluation of the image quality and the consequent reduction of dose in the patient. The developed simulator object for equipment evaluation of conventional radiology had been revealed efficient in the evaluation of the quality of the fluoroscopic image, as well as in the determination of the entrance surface air kerma rate of the normal adult patient skin. The evaluation of the parameters technician in the diverse ways of exposition of digital fluoroscopy equipment allowed establishing the pointers of improve dose for improve the quality of the images in interventionist procedures. This makes possible to reduce the dose in the patient keeping the quality of the image and, without loss of diagnostic information.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Mestrado.pdf (1.50 Mbytes)
Fecha de Publicación
2009-03-24
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.