Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.59.2012.tde-16042012-144526
Document
Auteur
Nom complet
Leonardo Lira do Amaral
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
Ribeirão Preto, 2012
Directeur
Jury
Ghilardi Netto, Thomaz (Président)
Nicolucci, Patrícia
Rodrigues, Laura Natal
Titre en portugais
Programa de controle da qualidade dosimétrico, validado com auxílio de filme radiocrômico, aplicado à radioterapia estereotáxica
Mots-clés en portugais
Dosimetria in vivo
Filme Radiocrômico
Programa de Controle da Qualidade em Radioterapia
Radioterapia
Resumé en portugais
A Radioterapia de lesões cerebrais próximas a estruturas críticas necessitam de uma alta precisão na localização e na dose. O rigor na liberação da dose deve ser acompanhado por um preciso controle da qualidade nos aparelhos que envolvam a prática. O comissionamento do sistema de planejamento consiste em averiguar e confirmar os cálculos realizados pelo sistema. Porém, mesmo com todo controle da qualidade no comissionamento, existem vários aspectos que podem influenciar na administração da dose no volume alvo, o que exige a necessidade de se fazer uma avaliação final, no ato do tratamento, in vivo. O objetivo deste trabalho é desenvolver uma técnica de dosimetria in vivo como parte de um programa de controle da qualidade em radioterapia estereotáxica. Na técnica de dosimetria in vivo, utilizaram-se segmentos de filme radiocrômico, com dimensões de 1x1 cm2, acoplados na área externa ao colimador formado por micro-lâminas, Moduleaf. Estes filmes foram inseridos na região central do feixe. Os filmes foram irradiados e calibrados para obtenção dos fatores campos, na configuração da técnica. Com estes dados foi elaborado um programa computacional, o qual calcula a densidade relativa que um filme deve adquirir quando submetido a uma exposição nesta configuração. Como a técnica de dosimetria in vivo usa os dados do TPS, validaram-se alguns parâmetros do comissionamento do TPS. Complementando o estudo da dosimetria in vivo foram avaliados cinco planos não co-planares, sendo o primeiro com 15 campos e os outros com 25 campos. Antes de iniciar o procedimento o segmento de filme era acoplado ao aparelho e após a execução do tratamento a densidade ótica era avaliada e comparada com a calculada pelo programa desenvolvido. No comissionamento, todas as dosimetrias relativas apresentaram diferenças percentuais menores que 2%, quando comparados os resultados medidos com os calculados pelo sistema de planejamento. No desenvolvimento da técnica de dosimetria in vivo, a diferença percentual média da verificação dosimétrica, no momento da irradiação, comparado com a calculada pela planilha foi de 1,5%, enquanto que a dosimetria absoluta aplicada ao controle da qualidade convencional foi aprovada com diferença percentual média de 2,5% e a função gama média encontrada foi de 97,9% dos pontos aprovados com critério de aceitação %=2% e D=2 mm. Logo, todos os dados estão em concordância com os limites estabelecidos pelo TRS-430. Desta forma, conclui-se que foi desenvolvida uma técnica de dosimetria in vivo como parte de um programa de controle da qualidade em radioterapia estereotáxica com filme radiocrômico, já que foram confirmados os parâmetros básicos do comissionamento do sistema de planejamento e a técnica foi validada com o controle de qualidade convencional nos cinco planos analisados.
Titre en anglais
Dosimetric quality assurance with the help of the radiochromic film, applied to stereotactic radiotherapy.
Mots-clés en anglais
in vivo dosimetry.
quality assurance
radiochromic film
Radiotherapy
Resumé en anglais
Radiation therapy of brain lesions near critical structures requires a highly accurate location and dose. The accuracy in dose delivery should be accompanied by an accurate quality control in devices involving the practice. The commissioning of the planning system is to ascertain and confirm the calculations performed by the system, but even with all quality control in the commissioning, there are several aspects that may influence the dosing the target volume, which necessitates the need to make a final assessment at the time of treatment, in vivo. The objective of this work is to develop a technique for in vivo dosimetry as part of quality assurance in stereotactic radiotherapy. In vivo dosimetry technique, we used segments of film radiocrômico, with dimensions of 1x1 cm2, coupled to the external area formed by the micro-collimator blades, Moduleaf. These films were inserted in the central region of the beam. The films were irradiated and calibrated to obtain factors of fields in the configuration of the technique. With these data we designed a computer program which calculates the relative density of a film must acquire when subjected to an exposure in this setting. As the technique of in vivo dosimetry using data from the TPS, validated parameters are the commissioning of the TPS. Complementing the study of in vivo dosimetry were evaluated five non-coplanar plans, the first with 15 fields and the other with 25 fields. Before starting the procedure, the film segment was attached to the unit and after the treatment is the optical density was measured and compared with those calculated by the program developed. At commissioning, all presented on dosimetry percentage differences less than 2%, when comparing the measured results with those calculated by the planning system. In developing the technique of in vivo dosimetry, the mean percent difference dosimetry verification at the time of irradiation compared with the calculated by the sheet was 1.5%, while the absolute dosimetry applied to the conventional quality control has been approved as mean percent difference 2.5% and the gamma function mean was 97.9% of the points agreed with the acceptance criterion % = 2% and D = 2 mm. Therefore, all data are in agreement with the limits set by TRS-430. Thus, we conclude that we have developed a technique for in vivo dosimetry as part of a quality assurance in stereotactic radiotherapy radiocrômico film, since some parameters were confirmed to commissioning the planning system and the technique was validated with control quality standard in five plans analyzed.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2012-06-15
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.