Title in Portuguese
Estudo e desenvolvimento de dosímetros luminescentes opticamente estimulados de a-Al2O3 com nanoestruturas de carbono e óxidos terras-raras para aplicação em radioterapia
Keywords in Portuguese
Al2O3
carbono grafite
Dosimetria OSL
nanoestruturas de carbono
Radioterapia.
terras-raras
Abstract in Portuguese
A alta sensibilidade e a reprodutibilidade de sinal são características de dosímetros luminescentes opticamente estimulados, que instigam seu uso no controle da qualidade e dosimetria in vivo em Radioterapia. Nesse trabalho, amostras de Al2O3, Al2O3 (1%C) e Al2O3 com diferentes concentrações de terras-raras e de nanoestruturas de carbono foram estudadas para aplicação em Radioterapia. As amostras foram obtidas pelo processo manual de mistura e maceração e inseridas em cápsulas transparentes, com exceção do material Al2O3 (1%C), que foi obtido na forma de pastilhas. Os sinais OSL para todos os materiais testados apresentaram crescimento de resposta com a dose de radiação. A concentração de 0,02% apresentou maior intensidade de sinal para todos os materiais inseridos na alumina, com exceção do carbono grafite (1%). Os materiais foram caracterizados por diferentes técnicas, para análises espectroscópicas e estruturais. A sensibilidade OSL do material Al2O3 (1%C) foi investigada para feixes de elétrons de 5 a 12 MeV, para feixe de fótons de 6 MV e para uma fonte de Irídio-192. Diferentes características dosimétricas OSL dos materiais Al2O3 (0,02%Eu), Al2O3 (0,02%Tb), Al2O3 (0,02%ntC), Al2O3 (0,02%npC) foram investigadas para feixes de fótons de 50 kVp, de 6 MV e para uma fonte de Irídio-192, em um amplo intervalo de doses. Todos os materiais apresentaram reprodutibilidade de sinal OSL (desvios menores que 1%) e homogeneidade de resposta em grupo de até ±95% (conferida para o dosímetro de Al2O3 (0,02%ntC). Todos os materiais estudados apresentaram sensibilidade satisfatória para doses de interesse clínico. Amostras OSL de Al2O3 (1% C), Al2O3 (0,02%ntC) e Al2O3 (0,02%npC) foram utilizadas para um estudo piloto de dosimetria in vivo em Radioterapia. A dosimetria in vivo sugere o possível uso dos dosímetros na rotina clínica em Radioterapia, através do estabelecimento de um protocolo de medidas, calibrações para cada feixe e energia utilizada e controle do sistema de planejamento utilizado para os cálculos de dose.
Title in English
Study and development of optically stimulated luminescent dosimeters of -Al2O3 with carbon nanostructures and rare-earth oxides for use in Radiotherapy
Keywords in English
Al2O3
carbono nanoestructures
earth-rares
graphte carbon
OSL Dosimetry
Radiotherapy
Abstract in English
High sensitivity and signal reproducibility are characteristics of optically stimulated luminescent dosimeter, instigating it use in quality control and in vivo dosimetry in Radiotherapy. In this work, samples of Al2O3, Al2O3 (1%C) and Al2O3 with different concentrations of rare-earth and carbon nanostructures, have been studied and developed for application in Radiotherapy. Samples were obtained by manual mixing and maceration and inserted into transparent capsules, except those of Al2O3 (1%C) which were produced as pellets. The OSL signals for all tested materials showed increasing response with radiation dose. The 0.02% concentration showed the highest signal intensity for all materials inserted into alumina, with exception to carbon graphite (1%). Materials were characterized by various techniques, through spectroscopic and structural analysis. The OSL sensitivity of Al2O3 (1%C) was investigated for electron beams from 5 to 12 MeV, 6 MV photons and an Iridium-192 source. Different dosimetric characteristics of some OSL materials Al2O3 (0.02%Eu), Al2O3 (0.02%Tb), Al2O3 (0.02%ntC), Al2O3 (0.02% npC) were investigated for photons beams of 50 kVp and 6 MV and a source of Iridium-192, in a wide dose range. All materials showed reproducibility of OSL signal (deviations less than 1%) and response homogeneity up to ± 95% (conferred for the dosimeter Al2O3 (0.02%ntC). All studied materials showed satisfactory sensitivity. Samples of Al2O3 (1%C), Al2O3 (0.02%ntC) and Al2O3 (0.02% npC) were used in a in vivo dosimetry pilot study in radiotherapy. The in vivo dosimetry performed with the developed OSL dosimeters suggest its possible use in clinical practice in Radiotherapy by establishing a measure protocol, calibrations for each beam and energy used and control of the planning system used for dose calculations.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2016-12-05
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.