Disertación de Maestría
DOI
https://doi.org/10.11606/D.59.2010.tde-04032011-155301
Documento
Autor
Nombre completo
Michely Cristina da Silveira
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
Ribeirão Preto, 2010
Director
Tribunal
Almeida, Adelaide de (Presidente)
Andrade, Jurandyr Moreira de
Bruni, Aline Thais
Título en portugués
Distribuição da dose absorvida no tratamento da micose fungóide através da dosimetria Fricke Xilenol Gel
Palabras clave en portugués
Dosimetria e Dosímetro Fricke Xilenol Gel
Radioterapia
Resumen en portugués
A radioterapia utiliza a radiação ionizante para a destruição de células tumorais. O controle da dose absorvida de um tipo específico de radiação aplicada a um volume alvo (tumor) é feito através da dosimetria, (uso de sensores de radiação), que neste trabalho foram utilizados os dosímetros químicos Fricke e filme. Dentre os diversos tipos de câncer, indicados para tratamento, utilizando a radiação ionizante, têm-se também a neoplasia Micose Fungóide (MF), linfoma que se alastra na superfície e em profundidade na pele para o qual elétrons com alta energia são utilizados para seu tratamento, devido aos seus alcances. Neste trabalho o Fricke Xilenol Gel (FXG) foi utilizado para o planejamento do tratamento da MF, para averiguar a distribuição da dose absorvida proveniente das interações dos elétrons com os tecidos. Para tal, foram construídas cubetas especiais em acrílico (tomos) conformacionadas aos contornos do crânio e do abdômen, que uma vez preenchidas com o FXG serviram para avaliar as doses absorvidas num objeto simulador antropomórfico, o Rando Phantom. O simulador, completado com cubetas e filmes radiocrômicos, este último também conformacionado aos dois contornos de interesse, foi submetido à técnica de Stanford para irradiação do corpo total com elétrons de 6 MeV. Imagens CCD e de densidade radiográfica foram feitas e posteriormente avaliadas, através de perfis horizontais e verticais passando por seus centros. Estes a seguir foram analisados por programas computacionais: um desenvolvido no Matlab para as imagens do FXG e o outro pelo programa já conhecido, ImageJ, para as imagens do filme. Das medidas, pode-se inferir que o dosímetro FXG se presta, similarmente ao filme, para as avaliações da dose absorvida devida aos elétrons na superfície do paciente e no seu interior, devida à contaminação por raios-X.
Título en inglés
Absorbed dose distribution in the treatment of Mycosis Fungoides using Fricke Xylenol Gel dosimetry
Palabras clave en inglés
Dosimetry
Fricke Xilenol Gel Dosimete.
Radioterapy
Resumen en inglés
Radioterapy uses ionizing radiation to destroy tumor cells. The absorbed dose control in a target volume is done through dosimetry, using radiation sensors, being the Fricke and film used in this study. Among several types of cancer indicated for treatment using ionizing radiation, there is also the Mycosis Fungoides (MF), lymphom that spreads on surface and depth in the skin, for which high-energy electrons are used for its treatment. In this work the Fricke Xylenol Gel (FXG) was used for MF treatment control, to obtain the absorbed dose distribution from electrons interaction with the tissues. For this scope cuvettes were manufactured of acrylic (slices), with the same contours of the skull and abdomen anthropomorphic simulator tomos. Once filled with the FXG, these cuvettes were used to infer the absorbed dose by the anthropomorphic simulator Rando Phantom. This simulator, completed with the cuvettes and radiochromic films, this last one also with the same contours similar to skull and abdomen were submitted to the Stanford technique, for 6 MeV electrons total body irradiation. CCD and radiographic density images were acquired and evaluated by horizontal and vertical profiles through theirs centers. These profiles were analyzed through a computer programs: one developed in Matlab for FXG images and the other by an already known program, ImageJ, for film images. From the results one can infer that the FXG dosimeter presents similarity with that of film, in the evaluation of the absorbed dose distribution on surface and also inside of the patient.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
MichelySilveira.pdf (1.28 Mbytes)
Fecha de Publicación
2011-06-20
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.