Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.59.2006.tde-02102007-112703
Document
Auteur
Nom complet
Ailton Massaiti Watanabe
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
Ribeirão Preto, 2006
Directeur
Jury
Bueno, Paulo Roberto (Président)
Faria, Ronaldo Censi
Mulato, Marcelo
Titre en portugais
Admitância/impedância eletroacústica aplicada ao estudo da formação de monocamadas automontadas e da afinidade proteínas/carboidratos
Mots-clés en portugais
admitância eletroacústica
biossensores
Microbalança de Cristal de Quartzo
monocamadas automontadas
Proteínas
Resumé en portugais
A técnica de Microbalança de Cristal de Quartzo (QCM) tem sido utilizada para a aplicação em biossensores. Esta técnica consiste no monitoramento da freqüência de oscilação do cristal de quartzo e a variação de freqüência está relacionada com a variação de massa na superfície do cristal por meio da equação de Sauerbrey. Porém, sabe-se que há discrepância entre a massa calculada pela equação de Sauerbrey e a massa real aderida na superfície do cristal de quartzo, sendo tal discrepância causada por fatores tais como a viscoelasticidade, o que é especialmente observado para macromoléculas e sistemas biológicos. Com a técnica de Admitância Eletroacústica (AE) é possível analisar a massa real devido à análise em separado dos parâmetros que compõem o circuito equivalente do cristal de quartzo e também inferir informações sobre a estrutura das monocamadas montadas sobre a superfície do cristal de quartzo. Neste trabalho foi utilizada técnica de AE para se potencializar e detalhar o estudo da formação de monocamadas automontadas e interações entre diferentes proteínas e carboidratos. Em muitos casos verificou-se que durante a formação de determinadas monocamadas, grande parte da variação de massa estava relacionada a interações fracas e não-específicas. Também durante a formação das monocamadas de tióis de cistamina, verificou-se que a estrutura da monocamada é extremamente dependente das concentrações iniciais de cistamina utilizadas para formar a monocamada. As interações entre proteínas e carboidratos foram realizadas após formação de multicamadas nas quais as proteínas eram imobilizadas para depois estudar-se, então, sua interação com carboidratos específicos. No caso da Jacalina, um tipo de lectina, observou-se interações bastante específicas com a fetuína e interações fracas com a alfa-lactose. A atividade da enzima GumH também foi testada com sucesso utilizando a técnica de AE.
Titre en anglais
Electroacoustic Admittance/ Impedance Applied to the Study of Self-Assembled Monolayers Formation and to the Study of Protein-Carbohydrate Affinity
Mots-clés en anglais
Biosensor
Electroacoustic Admittance
Protein
Quartz Crystal Microbalance
Self-Assembled Monolayers
Resumé en anglais
The Quartz Crystal Microbalance (QCM) has been largely used for biosensor applications. This technique consists on monitoring quartz crystal oscillation frequency. The relationship of quartz crystal frequency variation with mass deposited on the crystal surface is described by Sauerbrey´s equation. However, the relationship is not directly inferred and discrepancy exists caused mainly by viscoelasticity features of mass adsorbed on the quartz crystal surface. The viscoelastic influence is very common to be observed especially for macro-molecules and biological systems. By using Electroacoustic Admittance (EA) technique it is possible to calculate and separate the real mass deposited over the quartz crystal from other contributions. Therefore, the EA was used to infer about self-assembly monolayer structure being deposited over the quartz crystal. In this work EA technique was used to study details of self-assembled monolayer formation and specific interactions between different proteins and carbohydrate molecules. It was observed that during monolayer formation the majority of mass variation was related to non-specific interactions. For instance, regarding cystamine monolayer formation it was observed that the monolayer structure is dependent on the cystamine concentrations in the liquid medium. The study of interactions between proteins and carbohydrates was conducted by means of multilayers approach in which the proteins are immobilized previously to study the interaction between its active sites and carbohydrates. In the specific case of Jacalin, a type of lectin, it was observed a large formation of specific bonding between this protein and alpha-lactose or fetuin. The GumH enzyme activity was also tested successfully by using of EA technique.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
FINAL_DEFINITIVO.pdf (1.88 Mbytes)
Date de Publication
2008-04-04
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.