Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.54.1990.tde-06122013-150646
Document
Auteur
Nom complet
Pedro Luiz Frare Junior
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Carlos, 1990
Directeur
Jury
Tannus, Alberto (Président)
Baffa Filho, Oswaldo
Magon, Claudio Jose
Titre en portugais
Seleção de múltiplos planos em tomografia por ressonância magnética nuclear
Mots-clés en portugais
Imagens por Ressonancia Magnética
Pulsos de Radiofrequência Multibanda
Resumé en portugais
Apresentamos um estudo e a implementação de três técnicas destinadas a seleção de múltiplos planos, com objetivo de otimizar a tomografia por RMN. A primeira utiliza a excitação simultânea de múltiplos planos, sendo cada plano codificado com uma fase pré-determinada e a informação destes decodificada a posteriori através da combinação linear dos dados de n experimentos realizados. A segunda utiliza-se da excitação simultânea de múltiplos planos, como na técnica anterior, porém esses planos são adquiridos na presença de um gradiente de leitura oblíquo, que permite a obtenção simultânea dos sinais destes planos sem necessidade de pós-processamento. A terceira técnica usa a multiplexação de freqüências de excitação no tempo, isto é, diferentemente das teorias anteriores, excitam-se diferentes planos sucessivamente, durante o tempo de repetição de uma seqüência de aquisição, permitindo a aquisição de vários planos ao tempo de um único. A construção e o uso de um phantom destinado a caracterização do equipamento também é discutida. Esse phantom possibilita por exemplo: a determinação da largura do plano selecionado, o espaçamento entre os planos e o perfil destes, entre outros. Abordamos também, vários aspectos técnicos necessários a uma melhor performance do tomógrafo, tais como: circuitos de recepção e ressoadores. Apresentamos finalmente, uma discussão introdutória e os primeiros resultados experimentais já obtidos com a técnica de excitação adiabática com gradiente modulado (GMAX), utilizando bobinas de superfície
Titre en anglais
Simultaneous multislice slection in magnetic resonance tomography
Mots-clés en anglais
Magnetic Resonance Imaging
Multiband radiofrequency pulses
Resumé en anglais
We present the study and the implementation of three techniques HIS for the selection of multi-slice, aiming the optimization of the NMR tomography . The first technique uses the simultaneous excitation of multiple slices, being each plane encoded with a pre determined phase for a latter decodification of the information by the linear combination of n experiments. The second one makes use of the multi-slice simultaneous excitation, like the first one, but the slices are acquired in the presence of an oblique reading gradient , which provides us the multi-slice signal without any further computer processing. The third technique uses frequency multiplexed excitation, that is, different slices are successively excited during the repetition time of an acquisition sequence, making possible multi-slice acquisition at the same time of a single slice The construction and the use of a phantom for the equipment characterizations are discussed too. With this phantom we can determine the thickness of the selected slice, the spacing between the slices and the their shapes. Many technical aspects necessary for an improvement of the tomography performance, like reception circuits and resonators, are discussed. At last, we present a brief introduction to the gradient modulated adiabatic excitation (GMAX) technique and the first results ever obtained with it, using surface coils
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
PedroLuisFrareJrM.pdf (3.87 Mbytes)
Date de Publication
2013-12-09
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.