Master's Dissertation
DOI
https://doi.org/10.11606/D.54.1973.tde-19062015-170225
Document
Author
Full name
Christovam Mendonça Filho
Institute/School/College
Knowledge Area
Date of Defense
Published
São Carlos, 1973
Supervisor
Title in Portuguese
Espectrômetro de endor de faixa larga
Keywords in Portuguese
Não disponível
Abstract in Portuguese
Descreve-se a construção de um espectrômetro de ENDOR capaz de operar em uma única faixa de 0,5 a 75 MHz, utilizando um klystron apenas, com dereção homodina, sem exigir modificações fundamentais no espectrômetro de ressonância paramagnética eletrônica. O sinal de ENDOR pode ser detectado por modulação em amplitude de campo de rádio-frequência, ou por deteção direta do sinal de RPE, que é levado a um analisador, onde é feita a média dentre várias passagens, realçando a relação sinal/ruído. Como não há necessidade do uso de longas constantes de tempo, o sistema produz linhas de largura natural. O capítulo 1 apresenta os fundamentos das técnicas de RPE e ENDOR. No capítulo 2 é descrito o equipamento e os espectros obtidos estão no capítulo 3. No apêndice é feita uma breve discussão sob o ponto de vista dinâmico da experiência, relacionando as amplitudes das linhas de ENDOR com as condições experimentais, e se propõe um modelo para o seu mecanismo onde a relaxação cruzada adquire um papel de relevância
Title in English
Not available
Keywords in English
Not available
Abstract in English
One describe the construction of na ENDOR spectrometer operating from 0,5 to 75 MHz within a single band, with one klystron and homodine detection, and no fundamental changes on the electron spin resonance spectrometer. The ENDOR signal can be detected both by amplitude modulation of the frequency field, or direct detection of the ESR output, which is taken to a signal analyser, and the signal-to-noise ratio is raised by averaging rather than filtering, avoiding the use of long time constants, providing natural line widths. Chapter 1 present the basic principles of ESR and ENDOR techniques. In chapter 2 is described the experimental apparatus and the spectra obtained are in chapter 3. A brief discussion under dynamic viewpoint, relating the ENDOR line amplitudes with the experimental conditions is done in the Appendix, and an ENDOR mechanism in which there is a relevant presence of cross relaxation
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2015-06-19
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.