Master's Dissertation
DOI
https://doi.org/10.11606/D.5.2008.tde-25032009-091656
Document
Author
Full name
Tatiana de Arruda Ortiz
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2008
Supervisor
Committee
Tucci, Mauro Roberto (President)
Rodrigues, Marcus Vinicius Herbst
Volpe, Márcia Souza
Title in Portuguese
Influência do operador e do reanimador manual na manobra de hiperinsuflação manual: estudo em simulador do sistema respiratório
Keywords in Portuguese
Hiperinsuflação manual
Modalidades de fisioterapia
Terapia respiratória
Teste de materiais
Abstract in Portuguese
Introdução: A hiperinsuflação manual (HM) é uma técnica comum para facilitar a remoção de secreção traqueal de pacientes em UTI, e teoricamente, deveria ser realizada com uma insuflação lenta. Muitos fisioterapeutas não realizam a manobra com a insuflação lenta e os reanimadores manuais (RM) fabricados no Brasil possuem válvula de alívio de pressão, que não pode ser fechada na grande maioria desses dispositivos. Objetivos: Avaliar, em modelo mecânico, a influência da manobra de HM no volume corrente, picos de fluxo inspiratório e expiratório, e pressões alveolar e proximal por oito fisioterapeutas brasileiros, utilizando dois tipos de RM (com e sem válvula reguladora de pressão), de acordo com duas situações [conforme a prática clínica (PC) e após orientação de realizar a HM com a insuflação lenta, chamada de recomendada pela literatura (RL)], e em três cenários clínicos simulados (paciente normal, restritivo e obstrutivo). Resultados: Seis dos 8 fisioterapeutas realizaram a técnica com mais de duas insuflações; as pressões proximais geradas na situação RL foram menores devido aos menores picos de fluxo inspiratório. Os valores de pressões alveolares foram menores que 42,5 cmH2O (mediana = 13,9; intervalo interquartil:10,2-20,3) mesmo com altas pressões proximais (máximo 96,6 cmH2O, mediana 31,4; 19,2-44,8). Os volumes correntes foram menores do que os encontrados na literatura pesquisada (mediana = 514mL; 410-641). Os picos de fluxo inspiratórios (1,32; 0,92- 1,80) foram maiores que os expiratórios (0,88; 0,54-1,13) em quase todas as medidas. O pico de fluxo expiratório se correlacionou com o volume corrente: em cada cenário, os menores picos de fluxo expiratórios estavam correlacionados com baixos volumes correntes. Pressões, volumes e fluxos foram mais baixos com o RM com válvula reguladora de pressão. Conclusão: A manobra de HM foi realizada de forma diferente da preconizada na literatura; o RM com válvula reguladora de pressão gerou menores volumes, pressões e fluxos na maior parte dos cenários; as pressões proximais geradas não determinam, necessariamente, risco para o paciente, pois as pressões alveolares se mantiveram baixas
Title in English
Influences of operator and manual resuscitator on manual hyperinflation maneuver. A lung model study
Keywords in English
Manual hyperinflation
Materials testing
Physical therapy modalities
Respiratory therapy
Abstract in English
Background: Manual hyperinflation (MH) is a common technique used for removing pulmonary secretions in ICU and theoretically should be performed with a slow inflation. Many Brazilian respiratory therapists do not perform the maneuver with a slow inflation and manual resuscitators (MR) made in Brazil have pressure relief valve (PRV) which in many of them cannot be closed to perform MH. Objectives: evaluate, in a lung model, the influence on tidal volume, inspiratory and expiratory peak flow, proximal and alveolar pressures of MH performed by eight Brazilian respiratory therapists using two types of MR (with and without pressure relief valve), in two manners [like clinical practice (CP) and after orientation to perform MH with a slow inflation, named literature recommended (LR)] and in three clinical simulated scenarios (normal, restrictive and obstructive patient). Results: Six of 8 respiratory therapists performed MH with two or more inflation. After instruction proximal pressure generated was lower because of the slower inspiratory peak flow. The alveolar pressure values were lesser than 42.5 cmH2O (median = 13.9; interquartile range: 10.2-20.3), despite of high proximal pressure (max 96.6 cmH2O, median 31.4; 19.2-44.8). Inspiratory tidal volume were smaller than other reports (median=514mL; 410-641). Inspiratory peak flow (1.32; 0.92- 1.80) were higher than expiratory (0.88; 0.54-1.13) in almost all measurement. Expiratory peak flow was correlated with tidal volume: for each scenario, the low expiratory peak flow was mainly generated by small tidal volume. Pressures, volumes and flows were lower with MR that uses PRV. Conclusion: In this small sample of respiratory therapists, MH was done different that literature recommends; MR whit pressure limiting system generate lower tidal volume, pressures and peak pressures in most scenarios; proximal pressure generated not determine, necessarily, risk for patient, because alveolar pressure was maintained low
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
tatianaaortiz.pdf (1.19 Mbytes)
Publishing Date
2009-03-27
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.