• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Doctoral Thesis
DOI
https://doi.org/10.11606/T.5.2013.tde-16092013-162851
Document
Author
Full name
Sávia Christina Pereira Bueno
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2013
Supervisor
Committee
Costa, Roberto (President)
Fonseca, José Honório de Almeida Palma da
Scanavacca, Mauricio Ibrahim
Moreira, Luiz Felipe Pinho
Néto, Vicente Avila
Title in Portuguese
Estudo da segurança e da estabilidade do implante de cabo-eletrodo atrial direito pela via epidemiocárdiaca através do seio transverso em modelo animal suíno
Keywords in Portuguese
Eletrodos implantados
Marca-passo artificial
Modelos animais
Procedimentos cirúrgicos operatórios
Seios transversos
Suínos
Abstract in Portuguese
Introdução: Embora o implante de marcapasso atrioventricular pela via endovenosa seja considerado o estado da arte, situações clínicas e técnicas que podem impossibilitar o implante transvenoso de cabos-eletrodos têm sido cada vez mais frequentes, tornando-se necessária a proposição de abordagens cirúrgicas alternativas. Objetivos: O presente estudo visou o implante de marcapasso atrioventricular utilizando nova técnica para implante de cabo-eletrodo atrial no átrio direito, com o objetivo de avaliar no período intra-operatório e nos 30 primeiros dias de seguimento: a segurança e a reprodutibilidade do procedimento cirúrgico; a taxa de deslocamento dos cabos-eletrodos; as condições de estimulação e sensibilidade; as alterações morfológicas na cavidade pericárdica e as alterações histopatológicas epimiocárdicas. Métodos: Sob anestesia geral, foram operados 10 porcos adultos da raça Large White. Os cabos-eletrodos foram implantados, sob visão direta, no ventrículo esquerdo e no átrio direito, pelo seio transverso, através de toracotomia anterolateral esquerda. As condições de estimulação e de sensibilidade para os cabos-eletrodos atrial e ventricular, nas configurações unipolar e bipolar, foram avaliadas no intra-operatório, pós-operatório imediato, no 7º e 30º dias de pós-operatório. Ao final do estudo, os animais foram reoperados por toracotomia longitudinal mediana para observação das aderências pericárdicas e análise das condições histopatológicas da junção entre o cabo-eletrodo e o epimiocárdico. Resultados: Todos os animais permaneceram vivos até o último dia do estudo, não tendo ocorrido complicações intra-operatórias graves. As condições de estimulação e sensibilidade para os cabos-eletrodos atriais e ventriculares, nas configurações unipolar e bipolar, mantiveram-se estáveis ao longo do estudo e apresentaram comportamento semelhante. Notou-se aumento progressivo dos limiares atriais, variando de 0,50 ± 0,38 a 1,86 ± 1,31 volts nos períodos intra-operatório e 30o pós-operatório, respectivamente. Comportamento semelhante foi observado para os limiares ventriculares, que variaram de 0,43 ± 0,23 volts, no intra-operatório, a 1,22 ± 0,49 volts, no 30o pós-operatório. As medidas de impedância atrial e ventricular apresentaram uma discreta queda ao longo do tempo, sendo que a impedância atrial variou de 486,80 ± 126,35 a 385,0 ± 80,52 Ohms no período intra-operatório e 30o pós-operatório, respectivamente. A impedância ventricular variou de 700,40 ± 203,67 Ohms, no intra-operatório, a 409,30 ± 58,96 Ohms, no 30o pós-operatório. A sensibilidade, tanto em átrio quanto em ventrículo, mostrou-se estável a partir do pós-operatório imediato. A inspeção da cavidade pericárdica mostrou aderências em todos os animais e em todas as regiões avaliadas, não sendo observados derrame ou constrição pericárdica. A análise microscópica mostrou que o contato com os cabos-eletrodos provocou espessamento do epimiocárdio no átrio direito, ventrículo esquerdo e na artéria pulmonar. A cicatriz formada na junção entre o cabo-eletrodo e o epimiocárdio atrial foi semelhante à formada na região do implante dos cabos-eletrodos ventriculares. Conclusões: O implante do cabo-eletrodo atrial direito pelo seio transverso foi seguro e reprodutível. A efetividade do procedimento foi confirmada pelas condições estáveis de estimulação e de sensibilidade durante o período de seguimento pós-operatório. A única alteração morfológica encontrada na cavidade pericárdica foi a formação de aderências. Ocorreu adequada cicatrização na junção entre o cabo eletrodo e o epimiocárdio atrial
Title in English
Evaluation of safety and stability of right atrial lead implantation by epimyocardial access through the transverse sinus in a swine model
Keywords in English
Electrodes implanted
Models animal
Pacemaker artificial
Surgical procedures operative
Swine
Transverse sinuses
Abstract in English
Introduction: Although transvenous access for atrioventricular pacemaker implantation is considered the state of the art, clinical and technical situations that may impede transvenous leads implantation have become increasingly common, making it necessary the proposal of new surgical approaches. Objectives: The present study was design to perform the implant of atrioventricular pacemaker using a new technique for placement of the atrial lead in the right atrium, aiming to evaluate in the intraoperative period and during the first 30 days of follow-up: the safety and reproducibility of the surgical procedure; lead dislodgment rate; conditions of pacing and sensing; morphological changes in the pericardium, as well as, histopathological changes in the epimyocardium. Methods: A total of 10 Large White adult pigs underwent pacemaker implantation under general anesthesia. By using an anterolateral thoracotomy, leads were implanted under visual guidance in the left ventricle and in the right atrium through the transverse sinus. Pacing and sensing parameters, in unipolar and bipolar modes, were obtained during the intraoperative and immediate postoperative period and on the 7th and the 30th postoperative day. At the end of the study, all animals underwent reoperation by thoracotomy through a median longitudinal sternotomy for evaluation of pericardial adhesions and histopathological analysis of the junction between the lead and epimyocardial wall. Results: All animals were alive until the end of the study and there were no serious intraoperative complications. Pacing and sensing parameters for atrial and ventricular leads in both unipolar and bipolar modes remain stable throughout the study and showed similar performance. We observed a progressive increase in atrial thresholds, ranging from 0.50 ± 0.38 to 1.86 ± 1.31 volts, during the intraoperative and on the 30-day postoperative, respectively. Similar result was observed for ventricular thresholds, which ranged from 0.43 ± 0.23 volts, intraoperatively, to 1.22 ± 0.49 volts on the 30-day postoperative. Atrial and ventricular impedance measurements decreased slightly over time, and the atrial impedance ranged from 486.80 ± 126.35 to 385.0 ± 80.52 Ohms during the intraoperative and on the 30-day postoperative, respectively. Ventricular impedance ranged from 700.40 ± 203.67 Ohms, intraoperatively, to 409.30 ± 58.96 Ohms, on the 30-day postoperative. Both atrial and ventricular sensing measurements remain stable since the immediate postoperative period. Pericardial adhesions were found in all animals and in all studied regions. There was no noticeable pericardial effusion or myocardial constriction. Microscopic analysis showed that contact with the electrode had caused thickening of the right atrium, left ventricle and pulmonary artery. Similar scar formation was found at the interface between atrial lead and epimyocardial, such as ventricular lead and epimyocardial. Conclusion: Implantation of the right atrial lead through the transverse sinus was safe and reproducible. The effectiveness of the procedure was confirmed by stable conditions of pacing and sensing parameters throughout the postoperative follow-up. Pericardial adhesion was the only change found in the pericardial morphology. There was adequate scar formation between the interface of atrial lead and the epimyocardial
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2013-09-17
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.