• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Thèse de Doctorat
DOI
https://doi.org/10.11606/T.5.2013.tde-10052013-105804
Document
Auteur
Nom complet
Henrique Faria Ramos
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2013
Directeur
Jury
Brito Neto, Rubens Vuono de (Président)
Fávero, Mariana Lopes
Francesco, Renata Cantisani di
Hueb, Marcelo Miguel
Portinho, Fernando Sérgio de Melo
Titre en portugais
Avaliação da audição residual em candidatos  a implantes coclear atráves da resposta  auditiva de estado estável
Mots-clés en portugais
Implantes cocleares
Limiar auditivo
Perda auditiva neurossensorial
Potenciais evocados auditivos
Técnicas de diagnóstico em otologia
Resumé en portugais
INTRODUÇÃO: A identificação e preservação da audição residual em candidatos a implante coclear vêm assumindo maior importância clínica. A resposta auditiva de estado estável (RAEE) pode fornecer informação frequência-específica sobre o limiar auditivo em níveis de intensidade máximos de 120 dB NA, possibilitando a investigação da audição residual. Os objetivos deste estudo são avaliar a audição residual em candidatos a implante coclear, comparando os limiares eletrofisiológicos da RAEE com os limiares psicoacústicos da audiometria nas frequências de 500, 1 000, 2 000 e 4 000 Hz. MÉTODO: Foram avaliados 40 candidatos a implante coclear (80 orelhas) com perda auditiva neurossensorial severa a profunda bilateral. A audiometria foi realizada com o tom "warble" nas frequências de 500, 1 000, 2 000 e 4 000 Hz com intensidade máxima de estimulação de 120 dB NA. A RAEE foi obtida através da estimulação dicótica de uma frequência de tons contínuos sinusoidais modulados 100% em amplitude exponencial e 20% em frequência, nas frequências portadoras de 500, 1 000, 2 000 e 4 000 Hz, com estimulação máxima de 117, 120, 119 e 118 dB NA, respectivamente. RESULTADOS: Foram obtidos limiares mensuráveis em 62,5% de todas as frequências estudadas na audiometria tonal e em 63,1% na RAEE. A RAEE apresentou sensibilidade de 96% e especificidade de 91,6% na detecção da audição residual. As diferenças médias entre os limiares da audiometria tonal e da RAEE não apresentaram significância estatística em nenhuma das frequências. As correlações entre os limiares comportamentais e da RAEE foram significantes em todas as frequências avaliadas, sendo fortes em 500, 1 000 e 2 000 Hz e moderada em 4 000 Hz, com coeficiente de correlação de Pearson entre 0,65 e 0,81. Em 90% dos casos, os limiares da RAEE foram adquiridos nos limites de 10 dB dos limiares comportamentais. CONCLUSÕES: As correlações entre os limiares tonais e da RAEE foram significantes nas frequências de 500, 1 000, 2 000 e 4 000 Hz. A RAEE apresentou alta sensibilidade e especificidade na detecção da audição residual em candidatos a implante coclear, em comparação com a audiometria tonal
Titre en anglais
Evaluation of residual hearing in cochlear implants candidates using auditory steady-state response
Mots-clés en anglais
Auditory evoked potentials
Auditory thresholds
Cochlear implants
Otological diagnostic techniques
Sensorineural hearing loss
Resumé en anglais
INTRODUCTION: Identification and preservation of residual hearing in cochlear implantation are becoming more important lately. Auditory steadystate response (ASSR) can provide frequency-specific information regarding the auditory thresholds at maximum intensity levels of 120 dB HL, allowing investigation of residual hearing. The study objectives are to assess residual hearing in cochlear implant candidates by comparing the electrophysiological thresholds obtained in ASSR with psychoacoustic thresholds of audiometry at 500, 1 000, 2 000 and 4 000 Hz. METHOD: Forty cochlear implant candidates (80 ears) with bilateral severe-to-profound sensorineural hearing loss were studied. Warble-tone audiometry was performed at the frequencies 500, 1 000, 2 000 e 4 000 Hz, with stimuli up to 120 dB HL. ASSR was obtained with dichotic single-frequency stimulation of sinusoidal continuous tones modulated in exponential amplitude of 100% and in frequency of 20%, at the carrier frequencies of 500, 1 000, 2 000 and 4 000 Hz at maximum stimulation levels of 117, 120, 119 and 118 dB HL, respectively. RESULTS: Thresholds were obtained in 62,5% of all frequencies evaluated in warbletone audiometry and in 63,1% in the ASSR. ASSR showed sensitivity of 96% and specificity of 91,6% in the detection of residual hearing. The mean difference between the thresholds of behavioral audiometry and ASSR were not statistically significant in any of the frequencies. Strong correlations between behavioral and ASSR thresholds were observed in 500, 1 000 and 2 000 Hz and moderate in 4 000 Hz, with Pearson correlation coefficient between 0,65 and 0,81. In 90% of the occasions, ASSR thresholds were within 10 dB of behavioral thresholds. CONCLUSIONS: The correlations between behavioral and ASSR thresholds were significant at 500, 1 000, 2 000 and 4 000 Hz. ASSR showed high sensitivity and specificity in the detection of residual hearing in cochlear implant candidates, compared to warble-tone audiometry
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2013-05-10
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.