Tesis Doctoral
DOI
https://doi.org/10.11606/T.5.2016.tde-08042016-122000
Documento
Autor
Nombre completo
Luiz Lanziotti de Azevedo
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2015
Director
Tribunal
Silva, Erasmo Simao da (Presidente)
Bellen, Bonno Van
Miranda Junior, Fausto
Wolosker, Nelson
Zerati, Antonio Eduardo
Título en portugués
Injetor intravascular de CO2: estudo in vivo de viabilidade inicial do dispositivo e de qualidade de imagem
Palabras clave en portugués
Angiografia
Dióxido de carbono
Equipamentos médicos duráveis
Meios de contraste
Procedimentos endovasculares
Radiografia intervencionista
Resumen en portugués
INTRODUÇÃO: Durante procedimentos endovasculares, estão presentes os riscos relacionados ao uso dos contrastes iodados, tais como a nefropatia por contraste, uma vez que é fundamental o emprego de um meio de contraste para a obtenção das imagens radiológicas vasculares. A injeção intravascular de gás CO2 purificado é reconhecidamente uma alternativa relativamente mais segura ao contraste iodado, contudo, seu manuseio artesanal pode também trazer dificuldades técnicas e riscos aos pacientes. Para contemplar estas questões, foi desenvolvido o protótipo de um injetor intravascular de CO2 medicinal, microprocessado, dedicado à obtenção de imagens angiográficas. OBJETIVOS: Realizar os testes de viabilidade técnica inicial do protótipo em modelo in vivo. MÉTODOS: Realizar a angioplastia da artéria renal esquerda de 10 porcos, divididos em 2 grupos: Grupo 1 (n=5) injeção de contraste iodado, Grupo 2 (n=5) injeção de CO2 através do protótipo. Monitorização clínica e laboratorial dos animais no pré, intra e pós-operatório, com exames de função renal na véspera e 48h após os procedimentos e 3 gasometrias arteriais seriadas no intra-operatório. Observação clínica foi mantida por 48h no pós- operatório. RESULTADOS: Os procedimentos de angioplastia com CO2 foram realizados com sucesso técnico de 100%, sem necessidade de complementação com injeção de contraste iodado no Grupo 2. Não foram identificadas falhas no protótipo em funcionamento. Não foram identificadas alterações clínicas ou radiológicas sugestivas de contaminação por ar ambiente do sistema de CO2 e nem alterações laboratoriais nos animais. As imagens angiográficas obtidas no Grupo 2 foram consideradas, numa avaliação subjetiva, relativamente inferiores às imagens obtidas no Grupo 1. DISCUSSÃO: A qualidade inferior de imagem no Grupo 2 pode ser atribuída ao equipamento de fluoroscopia utilizado, com software desatualizado em relação aos equipamentos atuais, que incluem pré-configurações para angiografia com CO2; no entanto, ainda assim todos os procedimentos propostos no Grupo 2 foram realizados com sucesso técnico, o que nos leva a classificar as imagens deste grupo 2 como satisfatórias. O manuseio do aparelho mostrou-se ágil e eficiente, com a programação dos parâmetros sendo realizada com facilidade através do visor "touch screen", comparativamente superior ao método artesanal de injeção de CO2 com seringas em selo d'água. CONCLUSÕES: O protótipo do aparelho injetor intravascular de CO2 funcionou de forma adequada durante os testes e as imagens obtidas permitiram a compleição com sucesso dos procedimentos. Portanto, os resultados positivos obtidos sugerem que o equipamento é tecnicamente viável
Título en inglés
C02 intravascu/ar injector: in vivo ear/y feasibility study of the device and image quality assessment
Palabras clave en inglés
Angiography
Carbon dioxide
Contrast media
Durable medical equipment
Endovascular procedures
Interventional radiology
Resumen en inglés
INTRODUCTION: During endovascular procedures, several risks related to iodinated contrast media are taken, such as Contrast Induced Nephropathy, since these are fundamental to obtain radiological vascular images under fluoroscopy. Intravascular medical grade CO2 injection is a widely accepted and relatively safer alternative to iodinated contrast, but its handling and manual delivery system could as well bring risks to patients, such as room air injection. To address these problems, we developed a prototype for a micro processed intravascular medicinal CO2 injector, dedicated to help obtain angiographic images. OBJECTIVES: To perform the in vivo Early Feasibility Clinical Studies of the prototype. METHODS: Perform a left renal artery angioplasty of 10 pigs, divided in 2 groups: Group 1 (n=5) injection of iodinated contrast and Group 2 (n=5) injection of CO2 with the prototype. Clinical and laboratorial data were measured pre, peri and post-operatively, by pre-operative and 48h post- operative creatinine tests and 3 peri-operative serial arterial blood gas analysis. Animais were kept under clinical observation for 48h after procedures. RESUL TS: Angioplasty procedures had 100% technical success in both groups, without the need of extra imaging with iodinated contrast in Group 2. There were no technical failures with the prototype. There were no identifiable clinical or radiological signs of room air intravascular injections. Ali laboratorial measurements were under normal value ranges. Angiography images obtained in Group 2 were subjectively considered inferior to images from Group 1. DISCUSSION: The inferior image quality in Group 2 could be attributed to the employed fluoroscopy equipment, since its software is not up to date to current equipment, which includes a dedicated CO2 angiography preset. Nevertheless, ali procedures in Group 2 were successful, and images in this group were therefore considered satisfactory. Handling of the prototype proved to be user friendly, since ali injection parameters are performed through a touch screen panel, much superior from previous authors experience with manual saline submerged syringe filling before intravascular injection. CONCLUSIONS: The intravascular CO2 injector prototype worked properly during the tests and the obtained images allowed successful procedures without unexpected complications, in consonance with previous positive in vitro results. Therefore, the study results suggest the positive early technical feasibility of the device
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2016-05-12
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.